[Biomechanical evaluation of titanium mesh with anterior plate fixation or ilium autograft in anterior cervical decompression].

Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi

Department of Orthopaedics, the People's Hospital of Cuizhou Province,Guiyang Guizhou, 550002, P. R. China.

Published: November 2006

Objective: To evaluate the biomechanical characteristics of titanium mesh with anterior plate fixation or ilium autograft in anterior cervical decompression.

Methods: Six fresh cervical spine specimens (C3-7) of young cadaver were used in the biomechanical test. After C5, C5,6 and C4-6 were given vertebrectomy, ilium autograft and titanium mesh with anterior plate fixation were performed. Their stabilities of flexion, bilateral axial rotation, the lateral bending and the extension were tested. Intact cervical spine specimens served as control group.

Results: Ilium autograft improved the stability of the unstable cervical vertebrae and decreased the flexion, the lateral bending or the extension, showing a significant difference when compared with control group (P<0.05). Whereas, axial rotational motion was decreased insignificantly (P>0.05). Titanium mesh with anterior plate fixation improved the stability of the unstable spine and decreased the flexion,the bilateral axial rotation,the lateral bending or the extension, showing a significant difference when compared with control group (P<0.05).

Conclusion: The vertebrectomy and anterior cervical fusion by ilium autograft was the least stable construct of all modes tested,and the titanium mesh implantation is stable than the intact cervical sample.

Download full-text PDF

Source

Publication Analysis

Top Keywords

titanium mesh
16
mesh anterior
16
anterior plate
16
plate fixation
16
ilium autograft
16
lateral bending
12
bending extension
12
fixation ilium
8
autograft anterior
8
anterior cervical
8

Similar Publications

Alveolar ridge resorption following tooth loss poses a significant challenge for successful dental implant placement. In cases of severe atrophy, bone augmentation is required to restore sufficient bone volume. This technical note outlines a detailed, stepwise surgical protocol for horizontal and vertical alveolar ridge augmentation using customized titanium mesh.

View Article and Find Full Text PDF

Purpose: There are no universally established guidelines for material selection in orbital wall fracture reconstruction. With an increasing preference for permanent implants, this study aimed to compare the long-term clinical outcomes of three different non-resorbable materials in reconstructing isolated orbital floor fractures.

Design: A retrospective, interventional comparative study.

View Article and Find Full Text PDF

Metallic biomaterials are extensively used in orthopedics and dentistry, either as implants or coatings. In both cases, metal ions come into contact with surrounding tissues causing a particular cell response. Here, we present a biofabricated tissue model, consisting of a hydrogel reinforced with a melt electrowritten mesh, to study the effects of bound and released metal ions on surrounding cells embedded in a hydrogel matrix.

View Article and Find Full Text PDF

Background: Posterior vault distraction osteogenesis (PVDO) is an effective treatment for multisuture and syndromic craniosynostosis. It remains unclear how well the calvarium ossifies after PVDO, especially in older patients when they have plateaued in their cranial growth phase. The purpose of this study is to report outcomes associated with PVDO across a wide range of ages at our institution.

View Article and Find Full Text PDF

Evaluation of sticky bone in guided bone regeneration in the aesthetic area of the anterior teeth: a retrospective study.

Int J Oral Maxillofac Surg

January 2025

Stomatological Hospital of Chongqing Medical University, Chongqing, China; Chongqing Municipal Key Laboratory of Oral Biomedical Engineering of Higher Education, Chongqing, China; Chongqing Key Laboratory of Oral Diseases and Biomedical Sciences, Chongqing, China. Electronic address:

The retrospective study aimed to compare the space-maintaining effects of sticky bone (bone graft matrix enriched with injectable platelet-rich fibrin) and titanium mesh for bone augmentation in the aesthetic zone. Patients who underwent single implant placement and had type 2/4 alveolar bone defects (buccal bone wall loss is >50% of the expected implant length) were screened for inclusion in this study. The labial bone plate width was measured at 1, 3, and 5 mm below the apical implant platform on cone beam computed tomography images taken immediately and 6 months after surgery.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!