Background: Radio frequency ablation (RFA) has been accepted clinically as a useful local treatment for hepatocellular carcinoma (HCC). However, intrahepatic recurrence after RFA has been reported which might be attributable to increase in intra-tumor pressure during RFA. To reduce the pressure and ablation time, we developed a novel method of RFA, a multi-step method in which a LeVeen needle, an expansion-type electrode, is incrementally and stepwise expanded. We compared the maximal pressure during ablation and the total ablation time among the multi-step method, single-step method (a standard single-step full expansion with a LeVeen needle), and the method with a cool-tip electrode. Finally, we performed a preliminary comparison of the ablation times for these methods in HCC cases.
Results: A block of pig liver sealed in a rigid plastic case was used as a model of an HCC tumor with a capsule. The multi-step method with the LeVeen electrode resulted in the lowest pressure as compared with the single-step or cool-tip methods. There was no significant difference in the ablation time between the multi-step and cool-tip ablation methods, although the single-step methods had longer ablation times than the other ablation procedures. In HCC cases, the multi-step method had a significantly shorter ablation time than the single-step or cool-tip methods.
Conclusion: We demonstrated that the multi-step method was useful to reduce the ablation time and to suppress the increase in pressure. The multi-step method using a LeVeen needle may be a clinically applicable procedure for RFA.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1764754 | PMC |
http://dx.doi.org/10.1186/1476-5926-5-10 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!