Automated quality control in computed radiography.

Radiol Med

ASO San Giovanni Battista, C.so Bramante 88, I-10126, Torino (TO).

Published: December 2006

Purpose: The purpose of this paper is to describe the automation of quality control procedures on photo-stimulable imaging plates by means of an image-processing tool providing automatic reading of the images and automatic calculation of the quality parameters monitored.

Materials And Methods: Quality-control procedures were performed according to the main available guidelines. The quality assurance programme was applied to several Kodak and Philips devices in four radiological departments. The automatic image-processing tool was developed using public domain software (Java-based ImageJ software) and contains both reading and computation procedures.

Results: The quality checks and algorithms described were successfully applied, proving useful for identification of defective plates and for implementation of the quality assurance programme. The use of automation allowed significant savings in the time required for quality checks.

Conclusions: Completely automated image reading allows substantial economic and human resources savings, as it eliminates much of the transfer, reproduction, processing and filing procedures.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s11547-006-0113-5DOI Listing

Publication Analysis

Top Keywords

quality control
8
image-processing tool
8
quality assurance
8
assurance programme
8
quality
6
automated quality
4
control computed
4
computed radiography
4
radiography purpose
4
purpose purpose
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!