A pilot study to measure the compressive and tensile forces required to use retractable intramuscular safety syringes.

Am J Infect Control

College of Public Health, Department of Global Health, University of South Florida, 13201 Bruce B. Downs Boulevard, Tampa, FL 33612, USA.

Published: December 2006

Background: A pilot study was conducted at the Tampa Veterans Administration Patient Safety Center. The objective was to determine the forces required to operate retractable safety syringes to evaluate potential adverse occupational health and patient safety issues.

Methods: Four brands (100 each) of retractable syringes were tested, using a digital force gauge, in air and in a simulated patient material (SPM). Compressive and tensile forces were measured while activating the retraction mechanism and withdrawing saline into the syringe barrel, respectively.

Results: The mean compressive force was greater in SPM than in air in all 4 devices. There was a statistically significant compressive force difference between activation in air and SPM in devices 1 and 2 (P
Conclusion: This pilot study has implications for device selection, training for users, design issues for manufacturers, and patient safety, as well as potential for future needlestick and ergonomic injuries. We recommend replication of this study with a computer-controlled force testing apparatus, and by testing multiple needle and syringe sizes.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ajic.2006.04.212DOI Listing

Publication Analysis

Top Keywords

pilot study
8
compressive tensile
8
tensile forces
8
forces required
8
safety syringes
8
patient safety
8
compressive force
8
study measure
4
compressive
4
measure compressive
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!