Feasibility of calibrating elongated brachytherapy sources using a well-type ionization chamber.

Med Phys

Department of Radiation Medicine, University of Kentucky, 800 Rose Street, Lexington, Kentucky 40536, USA.

Published: November 2006

Recently, elongated brachytherapy sources (active length >1 cm) have become commercially available for interstitial prostate implants. These sources were introduced to improve the quality of brachytherapy procedures by eliminating the migration and seed bunching associated with loose seed-type implants. However, the inability to calibrate elongated brachytherapy sources with the Wide-Angle Free-Air Chamber (WAFAC) used by the National Institute of Standards and Technology (NIST) hinders the experimental determination of dosimetric parameters of these source types. In order to resolve this shortcoming, an interim solution has been introduced for calibration of elongated brachytherapy sources using a commercially available well-type ionization chamber. The feasibility of this procedure was examined by calibrating RadioCoil 103Pd sources with active lengths ranging from 1 to 7 cm.

Download full-text PDF

Source
http://dx.doi.org/10.1118/1.2358334DOI Listing

Publication Analysis

Top Keywords

elongated brachytherapy
16
brachytherapy sources
16
well-type ionization
8
ionization chamber
8
sources active
8
sources
6
brachytherapy
5
feasibility calibrating
4
elongated
4
calibrating elongated
4

Similar Publications

Background And Purpose: Elongation of life expectancy had led to marked increase in number of elderly patients with localized prostate cancer. However, the standard treatment for such patients is not well determined because of a high prevalence of comorbidities and slow growth of prostate cancer. The aim of this study is to examine the feasibility of radiotherapy for elderly patients aged ≥80 years.

View Article and Find Full Text PDF

Purpose: To present the clinical commissioning of a novel Pd directional brachytherapy device (CivaSheet) for intraoperative radiation therapy.

Methods And Materials: Clinical commissioning for the CivaSheet consisted of establishing: (1) source strength calibration capabilities, (2) experimental verification of TG-43 dosimetry parameters, (3) treatment planning system validation, and (4) departmental practice for dose specification and source ordering. Experimental verification was performed in water with radiochromic film calibrated with a 37 kVp X-ray beam.

View Article and Find Full Text PDF

[This corrects the article DOI: 10.1120/jacmp.v9i2.

View Article and Find Full Text PDF

Vulvar lymphangioma circumscriptum (LC) is a rare entity which may present as a painful, warty lesion. In contrast to the congenital form, which occurs in children, the acquired form arises in older adults and may be associated with infection, Crohn's disease, or prior pelvic/regional surgery. We present a case of acquired LC of the vulva in a 55-year-old woman who presented with a 3-4 year history of vulvar pain following chemotherapy, radiation, and brachytherapy for cervical cancer.

View Article and Find Full Text PDF

Purpose: To investigate quantitatively the deformation of the mammosite balloon, eccentricity of the source position and their effects on the dose delivered to the tumor lumpectomy site and critical structures.

Materials And Methods: The distances of the brachytherapy source to the surface of the mammosite balloon were measured in using radiographic images for fractions 1 to 10 for twelve patients. The dose at the balloon surface (nearly 680cGy) and prescription dose of 340cGy at 1cm from the balloon surface were calculated for the different fractions and their dependence on the balloon volume and source position were investigated.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!