Effect of 7.0 Tesla MRI on upper eyelid implants.

Ophthalmic Plast Reconstr Surg

Department of Otorhinolaryngology, Charité-Universitaetsmedizin Berlin, Charité Campus Mitte, Berlin, Germany.

Published: January 2007

To ensure the MRI compatibility of various eyelid implants in high-field MRI, 3 eyelid weights made of pure gold (99.99%), pure platinum (99.95%), and a platinum (97%)/iridium (3%) alloy were examined in vitro. Temperature changes, position changes, and imaging artifacts of the different implants were determined in a small-bore 7.0 Tesla MRI system. The 7.0 Tesla MRI system demonstrated that none of the eyelid implants carried a risk of heating or dislocation; therefore, these implants are MRI compatible up to a magnetic field strength of 7.0 Tesla.