Evaluation of the IRMA TRUpoint and i-STAT creatinine assays.

Clin Chim Acta

Department of Pathology, Tufts University School of Medicine, Baystate Health System, Springfield, MA 01199, United States.

Published: February 2007

Background: The i-STAT (Abbott Diagnostics, East Windsor, NJ) and IRMA TRUpoint (ITC, Edison, NJ) POCT analyzers were evaluated in an oncology center.

Methods: Precision and agreement with our core laboratory creatinine was judged by comparison of 50 consecutive chemotherapy patient results against the Roche rate-blanked Jaffe and enzymatic creatinine methods. Glomerular filtration rate (GFR) was estimated using the Cockroft-Gault (CG) calculation and Modification of Diet in Renal Disease Study (MDRD) equation.

Results: Precision varied from 1% (enzymatic)-6.1% (TRUpoint). Correlation was good (r>0.9948) with slopes within 5% of the Jaffe and enzymatic methods. Intercepts were <15.9 micromol/l (<0.18 mg/dl), and statistically significant bias (p<0.0025) was noted between the mean of patient specimens for i-STAT correlations to both the Jaffe and enzymatic laboratory creatinine methods. There was statistically significant concordance of estimated GFR between all methods, however, the agreement of estimated GFR to either the Jaffe or enzymatic creatinine laboratory methods was better for the TRUpoint (by either MDRD or CG estimation) and i-STAT (by MDRD equation) (Kappa>0.60) than the i-STAT (by CG estimation) (Kappa=0.41-0.60).

Conclusion: Small biases in the calibration of analytical creatinine methods can lead to differences in clinical concordance using estimated GFR. Selecting an optimal POCT method depends on the institution's current creatinine method and tolerance for analytical performance and clinical concordance.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2006.09.024DOI Listing

Publication Analysis

Top Keywords

irma trupoint
8
jaffe enzymatic
8
evaluation irma
4
trupoint i-stat
4
i-stat creatinine
4
creatinine assays
4
assays background
4
background i-stat
4
i-stat abbott
4
abbott diagnostics
4

Similar Publications

Point of care devices can assess electrolyte, blood gas, biochemical, and hematologic values in a critical care setting. Although these devices are commonly used in humans and companion mammals, few studies have assessed their use in avian species. This study compares electrolyte, hemoglobin (Hgb), hematocrit (Hct), acid-base, and venous blood gas parameters between the i-STAT and IRMA TruPoint blood gas analysis systems for 35 Quaker parrots.

View Article and Find Full Text PDF

Evaluation of the IRMA TRUpoint and i-STAT creatinine assays.

Clin Chim Acta

February 2007

Department of Pathology, Tufts University School of Medicine, Baystate Health System, Springfield, MA 01199, United States.

Background: The i-STAT (Abbott Diagnostics, East Windsor, NJ) and IRMA TRUpoint (ITC, Edison, NJ) POCT analyzers were evaluated in an oncology center.

Methods: Precision and agreement with our core laboratory creatinine was judged by comparison of 50 consecutive chemotherapy patient results against the Roche rate-blanked Jaffe and enzymatic creatinine methods. Glomerular filtration rate (GFR) was estimated using the Cockroft-Gault (CG) calculation and Modification of Diet in Renal Disease Study (MDRD) equation.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!