Objective: An important source of error in arterial spin labeling (ASL) is incomplete static tissue subtraction due to imperfect slice profiles. This effect can be reduced by saturating the spins in the imaging area prior to labeling. In this study, the use of optimized presaturation is compared with the use of optimized RF pulses for minimizing this error.
Materials And Methods: Different methods for optimizing presaturation were simulated by numerical solution of the Bloch equation. Presaturation was optimized by changing the number of presaturation pulses, their position in the pulse sequence and the area of the crusher gradients following each saturation pulse. It was also investigated whether the use of optimized presaturation could reduce the tag gap needed to ensure complete static tissue subtraction. Simulation results were verified using phantom and in vivo studies at 3T.
Results: In proximal inversion with control for off-resonance effects, optimized presaturation could reduce the necessary tag gap to 15% of the imaging slab for experiments using standard RF pulses, while c-FOCI RF pulses could reduce it to 11%. In flow-sensitive alternating inversion recovery, a single presaturation pulse could reduce the inversion width to 122% of the imaging slab and neither multiple presaturation pulses nor optimized RF pulses could reduce it further.
Conclusion: Optimized presaturation can reduce the necessary inversion width to the same level as if using optimized RF pulses and can, therefore, be used to optimize ASL sensitivity. Furthermore, optimized presaturation can reduce the B(1)-dependent sensitivity in static tissue subtraction.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.mri.2006.07.011 | DOI Listing |
Anal Chem
February 2025
Department of Biochemistry, Molecular Biology and Biophysics, University of Minnesota, Minneapolis, Minnesota 55455, United States.
H NMR spectroscopy has enabled the quantitative profiling of metabolites in various biofluids, emerging as a possible diagnostic tool for metabolic disorders and other diseases. To boost the signal-to-noise ratio and detect proton resonances near the water signal, current H NMR experiments require solvent suppression schemes (, presaturation, jump-and-return, WATERGATE, excitation sculpting, .).
View Article and Find Full Text PDFDrug Dev Ind Pharm
February 2025
Ramanbhai Patel College of Pharmacy, Charotar University of Science and Technology (CHARUSAT), Anand, India.
Background: Tavaborole (TAV), a benzoxaborole derivative, is an FDA-approved antifungal agent for treating onychomycosis, a common and persistent fungal infection of the toenails.
Objective: This study aimed to develop a robust stability-indicating HPTLC method to determine TAV in nanostructured lipid carriers (NLC) using a comprehensive approach that includes risk assessment, and Analytical Quality by Design.
Methods: The critical method parameters influencing the HPTLC results were screened using a Plackett-Burman screening design followed by its optimization using a central composite optimization design.
NMR Biomed
November 2024
Laboratory of Biomedical Imaging and Signal Processing, The University of Hong Kong, Hong Kong SAR, People's Republic of China.
We aim to explore the feasibility of head and neck time-of-flight (TOF) magnetic resonance angiography (MRA) at ultra-low-field (ULF). TOF MRA was conducted on a highly simplified 0.05 T MRI scanner with no radiofrequency (RF) and magnetic shielding.
View Article and Find Full Text PDFMagn Reson Med Sci
February 2024
Department of Radiology, National Center of Neurology and Psychiatry, Kodaira, Tokyo, Japan.
Purpose: Neuromelanin is visualized by optimizing the conditions of longitudinal relaxation (T1)-weighted imaging (T1WI). Although it was originally developed in 2D imaging, 3D imaging has been also reported, and T1WI sequences with magnetization transfer (MT) pulses are now widely used in 3D gradient echo (GRE) sequences. In this study, we assert that the use of spectral presaturation with inversion recovery (SPIR) may also be useful as an alternative to MT pulses, and we optimize SPIR and compare it with MT.
View Article and Find Full Text PDFMAGMA
April 2024
Institute of Radiology, University Hospital Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Erlangen, Germany.
Objective: To compensate subject-specific field inhomogeneities and enhance fat pre-saturation with a fast online individual spectral-spatial (SPSP) single-channel pulse design.
Methods: The RF shape is calculated online using subject-specific field maps and a predefined excitation k-space trajectory. Calculation acceleration options are explored to increase clinical viability.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!