Purpose: To compare the measurements of intraocular pressure (IOP) with Goldmann applanation tonometry (GAT) and dynamic contour tonometry (DCT) and the effects of central corneal thickness (CCT), corneal curvature, and level of IOP on these methods.

Design: Cross-sectional population-based study.

Participants: From the Los Angeles Latino Eye Study, 2157 participants of primarily Mexican ancestry.

Methods: Average GAT measurements were compared to DCT, and both were examined with respect to CCT (< or =500, 501-550, 551-600, >600 microns), corneal curvature (<42, 42-46, >46 diopters), and level of IOP (0-10, 11-20, >20 mmHg).

Main Outcome Measures: Mean GAT and DCT IOP levels were compared for the entire population, and then trends for the CCT, curvature, and IOP groupings were analyzed. The magnitude of the difference of GAT minus DCT was compared for these different strata, with special attention to a difference of +/- 3 mmHg or greater, which was defined as clinically significant.

Results: Mean IOP for the entire population by GAT was significantly lower (14.4+/-3.2 mmHg) compared with DCT (16.0+/-3.6; P<0.0001). Both GAT and DCT IOP levels were lowest for thin CCT and increased stepwise with increasing CCT, but this difference was more pronounced with GAT than DCT (P<0.0001 and P = 0.0012, respectively). The difference between GAT and DCT was largest for thin CCT and decreased for thicker CCT (P<0.0001). After adjusting for CCT, the corneal curvature affected IOP measured by DCT (P = 0.02) but not GAT (P = 0.3) such that mean DCT IOP increased with increasing corneal curvature. After adjusting for the CCT effect on IOP and stratifying by DCT IOP groups, the greatest difference between GAT and DCT was seen in the lowest IOP group (3.55+/-3.1), became negative in the intermediate group (-1.86+/-2.60), and was most negative in the highest IOP group (-3.88+/-3.3; P<0.0001).

Conclusions: Intraocular pressure measured by GAT was consistently lower when compared with DCT, and this difference was greatest with thinner CCT. Dynamic contour tonometry was also less affected by variations in CCT. Corneal curvature affected IOP measurements with DCT but not GAT, but this effect was less than the CCT effect on GAT. Goldmann applanation tonometry tended to underestimate IOP at higher levels and overestimate it at lower IOP levels when compared to DCT.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ophtha.2006.06.047DOI Listing

Publication Analysis

Top Keywords

corneal curvature
12
corneal thickness
8
intraocular pressure
8
goldmann applanation
8
applanation tonometry
8
dynamic contour
8
contour tonometry
8
level iop
8
compared dct
8
entire population
8

Similar Publications

We have compared the biomechanical properties of human and porcine corneas using vibrational optical coherence tomography (VOCT). The elastic modulus of the cornea has been previously reported in the literature to vary from about several kPa to more than several GPa based on the results of different techniques. In addition, the formation of corneal cones near the central cornea in keratoconus has been observed in the clinic.

View Article and Find Full Text PDF

Background And Objective: The study examines the relationship between ocular rotations and cardiovascular functions through detailed biomechanical analysis. The study documents specific patterns of ocular movements and their synchronization with cardiovascular activity, highlighting significant correlations. These findings provide a basis for understanding the opto-biomechanical interplay between ocular and cardiovascular dynamics.

View Article and Find Full Text PDF

Objective: This study aimed to investigate the effects of menopause on ocular health and visual function.

Method: Sixty-two women (31 premenopausal and 31 postmenopausal) who visited Rasht Health Center were selected and matched based on their ages and compared in terms of their hormonal status, central corneal thickness (CCT), contrast sensitivity (CS), binocular vision variables and refractive error.

Results: CCT and corneal curvature were significantly thinner in postmenopausal women.

View Article and Find Full Text PDF

Corneal curvature (CC) is a crucial ocular optical system parameter, indicating the shape of the cornea expressed in radii. This study aims to investigate the relationship between CC, age, gender, refractive errors, and central corneal thickness (CCT) in Saudi Arabia, a field lacking in research. Data were gathered from Imam Medical Center in Riyadh and Dr.

View Article and Find Full Text PDF

Purpose: Little is known about the effects of ambient atmospheric pressure (AP) on intraocular pressure (IOP) under ordinary conditions. This study aimed to investigate the effects of AP on Goldmann applanation tonometer-measured IOP (GAT-IOP) in normal eyes under everyday atmospheric conditions adjusting for effects of possible confounding factors including other climatic factors.

Methods: Data obtained from 2,431 normal healthy eyes of 2,431 subjects (mean age: 56.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!