AI Article Synopsis

  • The study focuses on creating a disposable screen printed carbon electrode (SPCE) on PVC using various inks for conducting and insulating purposes.
  • The best electrode configuration was determined through calibration tests, with the fifth configuration providing the highest sensitivity.
  • A modified version of this electrode was developed into a biosensor with horseradish peroxidase and gold nanoparticles, showing excellent performance in detecting hydrogen peroxide over a long period and a wide concentration range.

Article Abstract

The design and development of a screen printed carbon electrode (SPCE) on a polyvinyl chloride substrate as a disposable sensor is described. Six configurations were designed on silk screen frames. The SPCEs were printed with four inks: silver ink as the conducting track, carbon ink as the working and counter electrodes, silver/silver chloride ink as the reference electrode and insulating ink as the insulator layer. Selection of the best configuration was done by comparing slopes from the calibration plots generated by the cyclic voltammograms at 10, 20 and 30 mM K(3)Fe(CN)(6) for each configuration. The electrodes with similar configurations gave similar slopes. The 5th configuration was the best electrode that gave the highest slope. Modifying the best SPCE configuration for use as a biosensor, horseradish peroxidase (HRP) was selected as a biomaterial bound with gold nanoparticles (AuNP) in the matrix of chitosan (HRP/AuNP/CHIT). Biosensors of HRP/SPCE, HRP/CHIT/SPCE and HRP/AuNP/CHIT/SPCE were used in the amperometric detection of H(2)O(2) in a solution of 0.1M citrate buffer, pH 6.5, by applying a potential of -0.4V at the working electrode. All the biosensors showed an immediate response to H(2)O(2). The effect of HRP/AuNP incorporated with CHIT (HRP/AuNP/CHIT/SPCE) yielded the highest performance. The amperometric response of HRP/AuNP/CHIT/SPCE retained over 95% of the initial current of the 1st day up to 30 days of storage at 4 degrees C. The biosensor showed a linear range of 0.01-11.3mM H(2)O(2), with a detection limit of 0.65 microM H(2)O(2) (S/N=3). The low detection limit, long storage life and wide linear range of this biosensor make it advantageous in many applications, including bioreactors and biosensors.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.bios.2006.09.011DOI Listing

Publication Analysis

Top Keywords

design development
8
screen printed
8
printed carbon
8
carbon electrode
8
horseradish peroxidase
8
bound gold
8
gold nanoparticles
8
matrix chitosan
8
linear range
8
detection limit
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!