A combined technique with laser irradiation is suggested to control spark discharge for analytical use, having a unique feature that firing points of the spark discharge can be fixed by laser irradiation. Because the spark discharge easily initiates at particular surface sites, such as non-metallic inclusions, called selective discharge, the concentration of some elements sometimes deviates from their average one in spark discharge optical emission spectrometry. Therefore, stabilization of firing points on a sample surface could improve the analytical precision.