A 71-year-old woman with Turner's syndrome underwent pacemaker implantation for complete atrio-ventricular block. During the procedure, the persistence of left sided superior vena cava (LSVC) was observed such that the lead, through the coronary sinus, reached the right atrium. By use of stylets, we could drive the lead against the lateral atrial wall and curve it through the tricuspid valve into the right ventricle. The tip reached an apical stable position, obtaining proper stimulation values. Moreover, the VDD dipole was positioned against high lateral atrial wall, adequately sensing the atrial potential. So, we could obtain an atrial synchronous ventricular pacing with only one VDD lead.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8159.2006.00513.xDOI Listing

Publication Analysis

Top Keywords

vdd lead
8
superior vena
8
vena cava
8
turner's syndrome
8
lateral atrial
8
atrial wall
8
atrio-ventricular synchronization
4
synchronization single
4
single vdd
4
lead
4

Similar Publications

Background: An atrioventricular defibrillator system with a floating atrial dipole (VDD ICD) can provide atrial sensing by a single lead. Our aim was to compare the arrhythmia detection efficacy of VDD ICDs with conventional single- (VVI) and dual-chamber (DDD) defibrillators.

Methods: Data from consecutive patients undergoing ICD implantation were retrospectively analyzed.

View Article and Find Full Text PDF

Hepatic encephalopathy (HE) is a neuropsychiatric condition frequently associated with cirrhosis and portosystemic shunting (PSS). It imposes a significant clinical and economic burden, with increasing attention toward identifying modifiable factors that could improve outcomes. Emerging evidence suggests that vitamin D deficiency (VDD), prevalent in patients with cirrhosis, may contribute to the development and severity of HE.

View Article and Find Full Text PDF
Article Synopsis
  • - Vitamin D deficiency (VDD) is a major health concern affecting women and their babies, with evidence linking it to various complications during and after pregnancy, such as gestational diabetes and low birth weight.
  • - A review of literature from Europe over the past decade identified strong associations between VDD and adverse maternal and fetal health outcomes, including increased risks of preeclampsia, preterm birth, asthma, and autism spectrum disorder.
  • - The review calls for standardized guidelines for vitamin D supplementation during pregnancy, recommending daily intakes of 400-2,000 IU, and emphasizes the need for future research on consistent assessment methods and preventative strategies in prenatal care.
View Article and Find Full Text PDF
Article Synopsis
  • Conduction system pacing (CSP) is being explored as a new option compared to cardiac resynchronization therapy (CRT), especially with the use of implantable cardiac defibrillator (ICD) leads.
  • A study implanted a specific ICD lead in a swine model to test the effectiveness of a "single-lead" CRT-D setup.
  • The results showed that CSP was successfully achieved, indicating potential for fewer leads while maintaining essential functions such as defibrillation and atrial sensing, warranting further investigation.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!