Use of the Broselow tape may result in the underresuscitation of children.

Acad Emerg Med

National Flight Nurse Academy, Bolton School of Nursing, Case Western Reserve University, Cleveland, OH, 44106, USA.

Published: October 2006

Objectives: The purpose of this study was to determine the concordance of the Broselow tape with the measured heights and weights of a community-based population of children, especially in light of the increase in obesity in today's children.

Methods: The authors examined more than 7,500 children in a cross-sectional, descriptive study in two different cohorts of children to compare their actual weight with their predicted weight by a color-coded tape measure.

Results: In all patients, the percent agreement and kappa values of the Broselow color predicted by height versus the actual color by weight for the 2002A tape were 66.2% and 0.61, respectively. The concordance was best in infants, followed by school-age children, toddlers, and preschoolers (kappa = 0.66, 0.44, 0.39, and 0.39, respectively; percent agreement, 81.3%, 58.2%, 60.7%, and 64.0%, respectively). The tapes accurately predicted (within 10%) medication dosages for resuscitation in 55.3%-60.0% of the children. The number of children who were underdosed (by > or =10%) exceeded those who were overdosed (by > or =10%) by 2.5 to 4.4 times (p < 0.05). The tapes accurately predicted uncuffed endotracheal tube sizes when compared with age-based guidelines in 71% of the children, with undersizing (> or =0.5 mm) exceeding oversizing by threefold to fourfold (p < 0.05).

Conclusions: The Broselow tape color-coded system inaccurately predicted actual weight in one third of children. Caregivers need to take into consideration the accuracy of this device when estimating children's weight during the resuscitation of a child.

Download full-text PDF

Source
http://dx.doi.org/10.1197/j.aem.2006.06.042DOI Listing

Publication Analysis

Top Keywords

broselow tape
12
children
9
actual weight
8
percent agreement
8
tapes accurately
8
accurately predicted
8
weight
5
predicted
5
broselow
4
tape result
4

Similar Publications

Objective: To design a specific advanced life support (ALS) tape based on recent Indian multicenter height/length and weight data to accurately estimate the weight from the recumbent length.

Methods: We designed the new ALS tape by matching the median weights to median heights/lengths from the recently published Indian multicenter growth data, maintaining the same color codes as the Broselow tape. The accuracy of weight estimation for the newly designed ALS tape was validated and compared with the Broselow estimated weights at a tertiary care hospital.

View Article and Find Full Text PDF

Introduction: Determining the exact weight of children is a challenging task during emergency situations. Current guidelines recommend the use of length-based weight-estimating tapes. However, healthcare providers must either always carry the tapes or take time to locate them.

View Article and Find Full Text PDF

Comparing the performance of pediatric weight estimation methods.

Am J Emerg Med

August 2024

Division of Pediatric Emergency Medicine, Department of Pediatrics, Ann & Robert H. Lurie Children's Hospital of Chicago, 225 E Chicago Ave., Chicago, IL 60611, United States of America.

Background: Weight estimation is essential in the care of ill children when a weight cannot be obtained. This is particularly important for children with medical complexity, who are at higher risk for adverse drug events. Our objective was to compare the accuracy of different methods of weight estimation in children and stratify by the presence of medical complexity.

View Article and Find Full Text PDF

Objective: The use of pediatric length-based weight estimation tapes with precalculated drug doses is advocated by major Advanced Life Support organizations, but concerns have been raised on the accuracy of these systems. The objective of this systematic review was to collect, review, evaluate, and create a synthesis of the current literature to establish whether there is high-quality evidence for use of length-based tapes in accurate drug dose administration. A further objective was to compare these tapes with other dosing aids.

View Article and Find Full Text PDF

Unlabelled: Weight measurement is essential in the treatment of pediatric patients in emergencies, however, in cases of patient instability or limited resources, weight estimation becomes a plausible alternative. There are rapid estimation methods, although with performance discrepancies in different populations.

Objective: To compare the performance of the "Colombian Pediatric Tape" (CPT) and Broselow Tape (BT) in weight estimation in children.