A common strategy for measuring the real-ear response of the real-ear-to-coupler difference (RECD) in the pediatric population is to insert a probe-tube separately from the eartip. This strategy is at times difficult to implement while attempting to obtain the measurement from a young infant. An RECD probe-tube insertion technique that involves connecting the probe-tube to an eartip with plastic film for simultaneous insertion was examined on 30 infants. Repeated measurements were completed on each infant to obtain within-session test-retest reliability data. Probe-tube insertion depth was also examined across participants to provide a guideline for the infant population. Findings indicate that reliable RECD values can be obtained in infants when the probe-tube is extended approximately two to four millimeters (mm) beyond the eartip or 11 mm from the entrance to the ear canal. Clinical implications of this work are discussed.