Cyclic fatigue of different nickel-titanium endodontic rotary instruments.

Oral Surg Oral Med Oral Pathol Oral Radiol Endod

Faculty of Dentistry, University of Catania, Catania, Italy.

Published: October 2006

Objective: A comparative study of the fatigue resistance of rotary nickel-titanium endodontic instruments was performed with the aim of assessing the influence of both instrument design and surface treatment on flexural fracture.

Study Design: To evaluate fatigue resistance of different rotary instruments, a total of 120 instruments were tested; these came from different sources: ProFile, RaCe, K3, Hero, and Mtwo. To compare the effect of electro-polishing procedures on fatigue resistance, a group of RaCe instruments (which are normally electro-polished) without surface treatment was used. A scanning electron microscope (SEM) study for each instrument was performed before and after fatigue study to determine the mode of fracture and the aspect of tips and cross-sectional surface areas.

Results: ProFile instruments gave the best values for fatigue resistance. It was seen that for RaCe instruments the surface treatment reduces the presence of micro-cracks, surface debris, and machining damage.

Conclusions: This study shows that the instrument design often proves to be an important factor in the fatigue resistance of NiTi rotary instruments. In RaCe instruments the electro-polishing surface treatment increases the fracture-related fatigue resistance.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.tripleo.2005.12.012DOI Listing

Publication Analysis

Top Keywords

fatigue resistance
24
surface treatment
16
rotary instruments
12
race instruments
12
instruments
9
nickel-titanium endodontic
8
resistance rotary
8
instrument design
8
study instrument
8
fatigue
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!