Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 1034
Function: getPubMedXML
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3152
Function: GetPubMedArticleOutput_2016
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Objectives: The purpose of this study was to compare the biomechanical properties of posterolateral antiglide plating and lateral locked plating for fixation of displaced short oblique fractures of the fibula in osteoporotic bone.
Methods: Short oblique fractures of the distal fibula at the level of the syndesmosis were simulated with a fibular osteotomy and ligamentous sectioning in 18 paired fresh frozen ankles. The fractures were fixed with either a lateral locking plate with an independent lag screw or a posterolateral antiglide plate with a lag screw through the plate. The specimens were tested under a torsional load to failure. The torque to failure, angular rotation at failure, and construct stiffness of the two groups were compared.
Results: The torque to failure and construct stiffness were significantly greater on the side with the posterolateral antiglide plate than on the side with the the lateral locking plate (P = 0.01 and 0.005, respectively).
Conclusions: The posterolateral antiglide plate demonstrated improved biomechanical stability as compared to the lateral locking plate in osteoporotic bone. In situations where fixation needs to be optimized, use of an antiglide plate may be favored over a lateral locking plate construct.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/01.bot.0000245684.96775.82 | DOI Listing |