We report the utilization of a high power UV light-emitting diode for fluorescence detection (UV-LED-IF) in CE separations. CE-UV-LED-IF allows analysis of a range of environmentally and biologically important compounds, including PAHs and biogenic amines, including neurotransmitters, amino acids, proteins, and peptides, that have been derivatized with UV-excited fluorogenic labels, e.g., o-phthalic dicarboxaldehyde/beta-mercaptoethanol (OPA/beta-ME). The 365 nm UV-LED was used as a stable, low cost source for detection of UV-excited fluorescent compounds. UV-LED-IF was used with both zonal CE separations and MEKC. Native fluorescence detection of PAHs was accomplished with detection limits ranging from 10 nM to 1.3 microM. Detection limits for OPA/beta-ME-labeled glutamic acid and aspartic acid were 11 and 10 nM, respectively, for off-line labeling, and 47 and 47 nM, respectively, for on-line labeling, comparable to UV-laser-based systems. Analysis of OPA/beta-ME-labeled proteins and peptides was performed with 28 and 47 nM detection limits for BSA and myoglobin, respectively.