Purpose: This report describes a new image-guided radiotherapy (IGRT) technique using megavoltage cone-beam computed tomography (MV-CBCT) to treat paraspinous tumors in the presence of orthopedic hardware.

Methods And Materials: A patient with a resected paraspinous high-grade sarcoma was treated to 59.4 Gy with an IMRT plan. Daily MV-CBCT imaging was used to ensure accurate positioning. The displacement between MV-CBCT and planning CT images were determined daily and applied remotely to the treatment couch. The dose-volume histograms of the original and a hypothetical IMRT plan (shifted by the average daily setup errors) were compared to estimate the impact on dosimetry.

Results: The mean setup corrections in the lateral, longitudinal, and vertical directions were 3.6 mm (95% CI, 2.6-4.6 mm), 4.1 mm (95% CI, 3.2-5.0 mm), and 1.0 mm (95% CI, 0.6-1.3 mm), respectively. Without corrected positioning, the dose to 0.1 cc of the spinal cord increased by 9.4 Gy, and the doses to 95% of clinical target volumes 1 and 2 were reduced by 4 Gy and 4.8 Gy, respectively.

Conclusions: Megavoltage-CBCT provides a new alternative image-guided radiotherapy approach for treatment of paraspinous tumors in the presence of orthopedic hardware by providing 3D anatomic information in the treatment position, with clear imaging of metallic objects and without compromising soft-tissue information.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ijrobp.2006.05.038DOI Listing

Publication Analysis

Top Keywords

image-guided radiotherapy
12
paraspinous tumors
12
tumors presence
12
presence orthopedic
12
megavoltage cone-beam
8
cone-beam computed
8
computed tomography
8
treatment paraspinous
8
orthopedic hardware
8
imrt plan
8

Similar Publications

Background: Radiation-induced late fecal incontinence (LFI) is one of the most quality-of-life impairing symptoms in prostate cancer. We aimed to assess the impact of radiotherapy (RT) technique and dose-volume effects on LFI using a robust score.

Methods: We identified 409 patients who underwent curative intent using standard fractionated radiation therapy, 190 of them were finally included and analyzed.

View Article and Find Full Text PDF

Background: Preoperative chemoradiotherapy combined with total mesorectal excision (TME) is a standard treatment for locally advanced rectal cancer (LARC). However, lateral pelvic lymph nodes (LPLNs) are often inadequately treated with standard regimens. This study examines the treatment and postoperative outcomes in LARC patients receiving a simultaneous integrated boost (SIB) for LPLNs during long-course chemoradiotherapy.

View Article and Find Full Text PDF

Background And Purpose: Magnetic resonance imaging - linear accelerator (MRI-linac) systems permit imaging of tumours to guide treatment. Dynamic contrast enhanced (DCE)-MRI allows investigation of tumour perfusion. We assessed the feasibility of performing DCE-MRI on a 1.

View Article and Find Full Text PDF

Background And Purpose: A novel ring-gantry cone-beam computed tomography (CBCT) imaging system shows improved image quality compared to its conventional version, but its effect on autosegmentation is unknown. This study evaluates the impact of this high-performance CBCT on autosegmentation performance, inter-observer variability, contour correction times and delineation confidence, compared to the conventional CBCT.

Materials And Methods: Twenty prostate cancer patients were enrolled in this prospective clinical study.

View Article and Find Full Text PDF

Purpose: Minibeam radiotherapy (MBRT) uses small parallel beams of radiation to create a highly modulated dose pattern. The aim of this study is to develop an optical radioluminescence imaging (RLI) approach to perform real-time dose measurement for MBRT.

Methods: MBRT was delivered using an image-guided small animal irradiator equipped with a custom collimator.

View Article and Find Full Text PDF