Severity: Warning
Message: file_get_contents(https://...@pubfacts.com&api_key=b8daa3ad693db53b1410957c26c9a51b4908&a=1): Failed to open stream: HTTP request failed! HTTP/1.1 429 Too Many Requests
Filename: helpers/my_audit_helper.php
Line Number: 176
Backtrace:
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 176
Function: file_get_contents
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 250
Function: simplexml_load_file_from_url
File: /var/www/html/application/helpers/my_audit_helper.php
Line: 3122
Function: getPubMedXML
File: /var/www/html/application/controllers/Detail.php
Line: 575
Function: pubMedSearch_Global
File: /var/www/html/application/controllers/Detail.php
Line: 489
Function: pubMedGetRelatedKeyword
File: /var/www/html/index.php
Line: 316
Function: require_once
Unlabelled: The transthoracic impedance (T) and its variations may be estimated through the measurement of the electrical impedance between the can and the right ventricular coil of a defibrillation lead. This method may allow the monitoring of fluid overload before a heart failure attack. Aim of this study was to validate in vitro a method to calculate T in case of a standard bipolar pacing lead, by performing 3 measurements: standard unipolar impedance from the tip (Zuni-tip); unipolar impedance from the ring (Zuni-ring); standard bipolar impedance (Zbip). The formula we used is derived from the standard equivalent circuit of a pacing system: [Formula: see text] T represents the tissue impedance between the can and the electrodes of the lead. To validate the method we used a saline solution and 3 different pacing leads manufactured by Vitatron (Vitatron BV, Arnhem, The Netherlands): Impulse II (high impedance lead), Crystalline ActFix (screw-in lead), Brilliant S+ (VDD single-lead). The measured values of the saline solution impedance were compared to the values calculated through the formula.
Results: The calculated impedance of the solution, evaluated through the proposed formula, is reliable independently of the electrode used and highly correlated to the corresponding measured values (R>0.9).
Conclusion: Tissue impedance may be calculated from standard unipolar and bipolar impedance measurements with a standard bipolar pacing lead.