J Chromatogr A
Dalian Institute of Chemical Physics, Chinese Academy of Sciences, 457 Zhongshan Road, Dalian 116023, China.
Published: November 2006
An on-line immunoaffinity column with liquid chromatography/tandem mass spectrometry (IAC-LC-MS/MS) method for the determination of diuron in water matrices was described. This method used a sol-gel immunoaffinity column (20 mm x 4 mm I.D.) for on-line sample cleanup and enrichment, a monolithic analytical column (100 mm x 4.6 mm I.D.) for separation, and a triple quadrupole mass spectrometer for quantitation. The major challenges for the on-line set-up were discussed. The optimized on-line protocol was emphasized by the fact that low limit of quantitation (LOQ) of 1.0 ng/L was achieved with only 2.5-mL sample. In addition, a satisfactory accuracy ( approximately 90% of recovery) and precision (<6% of relative standard deviation) at 50 ng/L concentration were also obtained. Due to the ability of the sol-gel immunoaffinity column to eliminate matrix effect, the on-line IAC-LC-MS/MS analysis method can reliably determine diuron in wastewater treatment plant effluent sample.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.chroma.2006.08.007 | DOI Listing |
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!
© LitMetric 2025. All rights reserved.