A simple, rapid, and selective on-line method for the speciation and determination of Cr(III) and Cr(VI) in aqueous solutions by ion-pairing HPLC coupled with flame atomic absorption spectrometry (FAAS) is described. The composition of the mobile phase has been optimized for better separation. The effects of column temperature, volume of injection loop, fuel flow rate of FAAS, and nebulizer suction rate of FAAS have also been investigated. Separation is accomplished in almost 2.5 min on a 25 cm length C18 column at 40 degrees C. The selectivity of the method has been established by investigating the effect of interfering elements on chromium determination. The detection limit (3sigma) achieved by the method was calculated as 3.7 ng/mL for Cr(III) and 2.0 ng/mL for Cr(VI). The proposed method has been validated by analyzing certified reference material (BCR 544) and successfully applied to the analysis of drinking water and wastewater samples with a relative error below 6%.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1002/jssc.200600057 | DOI Listing |