This study sought to compare the cutting efficiency of different diamond burs on initial use as well as during repeated use, alternating with sterilization. Long, round-end, tapered diamond burs with similar diameter, profile, and diamond coarseness (125-150 microm grit) were used. A high-torque, high-speed electric handpiece (set at 200,000 rpm) was utilized with a coolant flow rate of 25 mL/min. Burs were tested under a constant load of 170 g while cuts were made on a machinable ceramic substrate block. Each bur was subjected to five consecutive cuts for 30 seconds of continuous operation and the cutting depths were measured. All burs performed similarly on the first cut. Cutting efficiencies for three of the bur groups decreased significantly after the first cycle; however, by the fifth cycle, all bur groups performed similarly without any significant differences (p > 0.05). A scanning electron microscope revealed significant crystal loss after each use.