Background: In mechanically ventilated patients, we investigated how positioning the heat and moisture exchanger (HME) at different places on the ventilator circuit affected inspiratory gas humidification.
Methods: Absolute humidity (AH) and temperature (TEMP) at the proximal end of endotracheal tube (ETT) were measured in ten mechanically ventilated patients. The HME was connected either directly proximal to the ETT (Site 1) or at before the circuit Y-piece (Site 2: distance from proximal end of ETT and Site 2 was about 19 cm) (Figure. 1). Two devices, Hygrobac S (Mallinckrodt Dar, Mirandola, Italy) and Thermovent HEPA (Smiths Medical International Ltd., Kent, UK) were tested. AH and TEMP were measured with a hygrometer (Moiscope, MERA Co., Ltd., Tokyo, Japan).
Results: Hygrobac S provided significantly higher AH and TEMP at both sites than Thermovent HEPA. Both Hygrobac S and with Thermovent HEPA provided significantly higher AH and TEMP when placed proximally to the ETT.
Conclusion: Although placement proximal to the ETT improved both AH and TEMP in both HMEs tested, one HME performed better in the distal position than the other HME in the proximal position. We conclude the both the type and placement of HME can make a significant difference in maintaining AH and TEMP during adult ventilation.
Download full-text PDF |
Source |
---|---|
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1562445 | PMC |
http://dx.doi.org/10.1186/1471-2466-6-19 | DOI Listing |
Pediatr Pulmonol
November 2024
Department of Community Medicine, Nagasaki University Graduate School of Biomedical Sciences, Nagasaki, Japan.
Introduction: Neonatal ventilators are managed by monitoring the inspiratory gas temperature with a probe placed outside the incubator, although the temperature decreases as the gases passe through the ventilator circuit extension tube and endotracheal tube (ETT). There are no established methods for measuring the internal temperature of the ETT. This bench study aimed to investigate the feasibility of using infrared thermography (IRT) to predict the internal temperature of the ETT.
View Article and Find Full Text PDFIran Endod J
January 2024
Department of Endodontics, School of Dentistry, Shahid Beheshti University of Medical Sciences, Tehran, Iran.
Introduction: Endodontic therapy has been regarded as a safe treatment option with highly promising results in preserving natural teeth. Numerous factors may influence the longevity of endodontically treated teeth (ETT), thus altering their success/survival rate. This study aimed to determine the survival and success rates of endodontically treated premolars and molars with different complex restorative/prosthodontic treatments.
View Article and Find Full Text PDFJ Clin Exp Dent
August 2024
Associate Professor, Department of Operative Dentistry, School of Dentistry, Shahed University, Tehran, Iran.
Background: This study aimed to compare the efficacy of resin composite for establishing a proper proximal contact in comparison to digital work flow Zirconia for restoration of endodontically treated teeth (ETT).
Material And Methods: Forty patients with posterior root canal treated teeth considering the inclusion and exclusion criteria were divided into two groups: half were restored by resin composite while the other received zirconia crown. Then, the proximal contact tightness (PCT) was measured via two methods: 1: VAS: the magnitude of felt PCT was recorded as a number between 0 to 10 (Visual Analogue Scale (VAS)).
Saudi J Anaesth
June 2024
Department of Anesthesiology and Pain Medicine, Nationwide Children's Hospital, Columbus, Ohio, USA.
Background: The use of cuffed endotracheal tubes (ETTs) has become the standard of care in pediatric practice. The rationale for the use of a cuffed ETT is to minimize pressure around the cricoid while providing an effective airway seal. However, safe care requires that the cuff lie distal to the cricoid ring following endotracheal intubation.
View Article and Find Full Text PDFJ Crit Care Med (Targu Mures)
April 2024
Jubilee Mission Medical College and Research Institute, Thrissur, Kerala, India.
Introduction: Airway ultrasound has been increasingly used in correct positioning of endotracheal tube. We hypothesize that a safe distance between endotracheal tube tip and carina can be achieved with the aid of ultrasound.
Aim Of The Study: Our primary objective was to determine whether ultrasound guided visualisation of proximal end of endotracheal tube cuff is better when compared to conventional method in optimal positioning of tube tip.
Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!