Effect of heat and moisture exchanger (HME) positioning on inspiratory gas humidification.

BMC Pulm Med

Department of Emergency and Critical Care Medicine, The University of Tokushima Graduate School, 3-18-15 Kuramoto, Tokushima, 770-8503, Japan.

Published: August 2006

Background: In mechanically ventilated patients, we investigated how positioning the heat and moisture exchanger (HME) at different places on the ventilator circuit affected inspiratory gas humidification.

Methods: Absolute humidity (AH) and temperature (TEMP) at the proximal end of endotracheal tube (ETT) were measured in ten mechanically ventilated patients. The HME was connected either directly proximal to the ETT (Site 1) or at before the circuit Y-piece (Site 2: distance from proximal end of ETT and Site 2 was about 19 cm) (Figure. 1). Two devices, Hygrobac S (Mallinckrodt Dar, Mirandola, Italy) and Thermovent HEPA (Smiths Medical International Ltd., Kent, UK) were tested. AH and TEMP were measured with a hygrometer (Moiscope, MERA Co., Ltd., Tokyo, Japan).

Results: Hygrobac S provided significantly higher AH and TEMP at both sites than Thermovent HEPA. Both Hygrobac S and with Thermovent HEPA provided significantly higher AH and TEMP when placed proximally to the ETT.

Conclusion: Although placement proximal to the ETT improved both AH and TEMP in both HMEs tested, one HME performed better in the distal position than the other HME in the proximal position. We conclude the both the type and placement of HME can make a significant difference in maintaining AH and TEMP during adult ventilation.

Download full-text PDF

Source
http://www.ncbi.nlm.nih.gov/pmc/articles/PMC1562445PMC
http://dx.doi.org/10.1186/1471-2466-6-19DOI Listing

Publication Analysis

Top Keywords

proximal ett
12
thermovent hepa
12
heat moisture
8
moisture exchanger
8
exchanger hme
8
inspiratory gas
8
mechanically ventilated
8
ventilated patients
8
ett site
8
provided higher
8

Similar Publications

Introduction: Neonatal ventilators are managed by monitoring the inspiratory gas temperature with a probe placed outside the incubator, although the temperature decreases as the gases passe through the ventilator circuit extension tube and endotracheal tube (ETT). There are no established methods for measuring the internal temperature of the ETT. This bench study aimed to investigate the feasibility of using infrared thermography (IRT) to predict the internal temperature of the ETT.

View Article and Find Full Text PDF

Introduction: Endodontic therapy has been regarded as a safe treatment option with highly promising results in preserving natural teeth. Numerous factors may influence the longevity of endodontically treated teeth (ETT), thus altering their success/survival rate. This study aimed to determine the survival and success rates of endodontically treated premolars and molars with different complex restorative/prosthodontic treatments.

View Article and Find Full Text PDF

Background: This study aimed to compare the efficacy of resin composite for establishing a proper proximal contact in comparison to digital work flow Zirconia for restoration of endodontically treated teeth (ETT).

Material And Methods: Forty patients with posterior root canal treated teeth considering the inclusion and exclusion criteria were divided into two groups: half were restored by resin composite while the other received zirconia crown. Then, the proximal contact tightness (PCT) was measured via two methods: 1: VAS: the magnitude of felt PCT was recorded as a number between 0 to 10 (Visual Analogue Scale (VAS)).

View Article and Find Full Text PDF

Background: The use of cuffed endotracheal tubes (ETTs) has become the standard of care in pediatric practice. The rationale for the use of a cuffed ETT is to minimize pressure around the cricoid while providing an effective airway seal. However, safe care requires that the cuff lie distal to the cricoid ring following endotracheal intubation.

View Article and Find Full Text PDF

Introduction: Airway ultrasound has been increasingly used in correct positioning of endotracheal tube. We hypothesize that a safe distance between endotracheal tube tip and carina can be achieved with the aid of ultrasound.

Aim Of The Study: Our primary objective was to determine whether ultrasound guided visualisation of proximal end of endotracheal tube cuff is better when compared to conventional method in optimal positioning of tube tip.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!