AI Article Synopsis

Article Abstract

Objectives: To develop a less expensive assay to calculate HIV-1 viral load for use in resource-limited countries.

Material And Method: An In-house One-tube-one-step Viral Load Assay (IOVA) was developed by using real-time PCR-based with TaqMan probe. Primers and probe were designed from the conserved region of sequences from all HIV subtypes. A standard curve was generated from reference virus in various dilutions. IOVA was applied on 105 HIV-positive and 25 HIV-negative samples and compared with the results from ROCHE AMPLICLOR.

Results: IOVA measured HIV RNA in the samples ranging from 125 to 2 x 10(6) copies/mL. The coefficient of variation of intra- and inter-assay ranged from 0.68% to 7.89%. The sensitivity, specificity, positive and negative predictive values were 92%, 100%, 100% and 79.5% respectively. The parallel quantitative analysis showed high correlation (r=0.95) between IOVA and AMPLICOR.

Conclusion: A new HIV-1 viral load assay was developed and validated. It was reliable and less expensive.

Download full-text PDF

Source

Publication Analysis

Top Keywords

viral load
12
taqman probe
8
hiv-1 viral
8
load assay
8
quantification hiv-1
4
hiv-1 rna
4
load
4
rna load
4
load one-tube-one-step
4
one-tube-one-step pcr
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!