Introduction: Many patients with implantable cardioverter defibrillators (ICDs) have older lead systems, which are usually not replaced at the time of pulse generator replacement unless a malfunction is noted. Therefore, optimization of defibrillation with these lead systems is clinically important. The objective of this prospective study was to determine if an active abdominal pulse generator (Can) affects chronic defibrillation thresholds (DFTs) with a dual-coil, transvenous ICD lead system.

Methods And Results: The study population consisted of 39 patients who presented for routine abdominal pulse generator replacement. Each patient underwent two assessments of DFT using a step-down protocol, with the order of testing randomized. The distal right ventricular (RV) coil was the anode for the first phase of the biphasic shocks. The proximal superior vena cava (SVC) coil was the cathode for the Lead Alone configuration (RV --> SVC). For the Active Can configuration, the SVC coil and Can were connected electrically as the cathode (RV --> SVC + Can). The Active Can configuration was associated with a significant decrease in shock impedance (39.5 +/- 5.8 Omega vs. 50.0 +/- 7.6 Omega, P < 0.01) and a significant increase in peak current (8.3 +/- 2.6 A vs. 7.2 +/- 2.4 A, P < 0.01). There was no significant difference in DFT energy (9.0 +/- 4.6 J vs. 9.8 +/- 5.2 J) or leading edge voltage (319 +/- 86 V vs. 315 +/- 83 V). An adequate safety margin for defibrillation (> or =10 J) was present in all patients with both shocking configurations.

Conclusion: DFTs are similar with the Active Can and Lead Alone configurations when a dual-coil, transvenous lead is used with a left abdominal pulse generator. Since most commercially available ICDs are only available with an active can, our data support the use of an active can device with this lead system for patients who present for routine pulse generator replacement.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8167.2006.00374.xDOI Listing

Publication Analysis

Top Keywords

pulse generator
24
abdominal pulse
16
dual-coil transvenous
12
generator replacement
12
active abdominal
8
defibrillation thresholds
8
transvenous icd
8
lead
8
icd lead
8
lead system
8

Similar Publications

Background: Despite the physiological advantages of positive end-expiratory pressure (PEEP), its optimal utilization during one-lung ventilation (OLV) remains uncertain. We aimed to investigate whether individualized PEEP titration by lung compliance is associated with a reduced risk of postoperative pulmonary complications during OLV.

Methods: We searched PubMed, Embase, and the Cochrane Central Register of Controlled Trials until April 1, 2024, to identify published randomized controlled trials that compared individualized PEEP titration by lung compliance with fixed PEEP during OLV.

View Article and Find Full Text PDF

Catheter-directed therapy for pulmonary embolism in pediatrics: a systematic review and meta-analysis.

Thromb J

January 2025

Pediatric Emergency Department, St. Christopher's Hopsoital for Children, Philadelphia, PA, USA.

Background: Acute pulmonary embolism (PE) is a serious and potentially fatal condition that is relatively rare in the pediatric population. In patients presenting with massive/submassive PE, catheter-directed Therapy (CDT) presents an emerging therapeutic modality by which PE can be managed.

Methods: Electronic databases were systematically searched through May 2024.

View Article and Find Full Text PDF

Background: Nemonoxacin is a new quinolone with an antibacterial efficacy against methicillin-resistant Staphylococcus aureus (MRSA). Certain sequence types (STs) have been emerging in Taiwan, including fluoroquinolone-resistant ST8/USA300. It's an urgent need to determine nemonoxacin susceptibility against ST8/USA300 and other emerging lineages, if any.

View Article and Find Full Text PDF

Background: Community-acquired pneumonia (CAP) was one of the most common causes of death in the European Union in 2017. Severity and mortality of CAP increase with age and an aging European population will require increased planning for prevention, control, and management of CAP. The purpose of this study was to provide an updated population-based estimate of the incidence of CAP requiring hospitalization in Northern Europe.

View Article and Find Full Text PDF

Many theories of time perception propose the existence of an internal pacemaker, and studies across behavioral, physiological, and neuroscience fields have explored this concept. Specifically, Spontaneous Motor Tempo (SMT), the most comfortable and natural tapping tempo for each individual, is thought to reflect this internal pacemaker's tempo. Changes in heart rate are also linked to time estimation, while Individual Alpha Frequency (IAF), the peak in the alpha range (8-13 Hz) observed in EEG, is reported to reflect the brain's temporal processing.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!