Numerical assessment of the impact of a flow wire on its velocity measurements.

Ultrasound Med Biol

Cardiovascular Mechanics and Biofluid Dynamics, Institute of Biomedical Technology, Ghent University, Ghent, Belgium.

Published: July 2006

Blood flow velocities can be measured using a Doppler flow wire. This numerical study evaluates the impact of a 0.014" flow wire on the measured frequencies in a straight artery with diameters of 3 mm and 4 mm, under steady and pulsatile flow conditions. Simulations were performed with the wires positioned differently in the artery (perfectly centred and at an offset of 0.5 mm from the wall) and with different types of wire (tilted and straight). Measurements were taken at range gates from 4 mm to 10 mm. During simulations using a 3-mm vessel under pulsatile flow conditions, the relative error between the measured and reference maximum frequency (occurring in absence of the wire) decreased from 17.7% to 11.6% (with a mean value of 14.9%). During simulations using an off-centre 1.5 degree tilted wire, the mean error was approximately 5%. Therefore, our study suggests that that a centrally positioned flow wire is unfavourable for measuring flow velocities.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.ultrasmedbio.2006.03.016DOI Listing

Publication Analysis

Top Keywords

flow wire
16
flow
8
flow velocities
8
pulsatile flow
8
flow conditions
8
wire
7
numerical assessment
4
assessment impact
4
impact flow
4
wire velocity
4

Similar Publications

Background: Severe vessel tortuosity may prevent a microcatheter from reaching a distal vessel. However, the Double-Wire Technique (DWT) may facilitate the procedure. The present study evaluated the feasibility and safety of guiding a 0.

View Article and Find Full Text PDF

Spontaneous Cargo Transport via Sliding Bubbles on a Superhydrophobic Wire.

Langmuir

December 2024

Zhejiang Provincial Key Laboratory of Flow Measurement Technology, China Jiliang University, Hangzhou 310018, China.

The transportation and carrying behavior of underwater bubbles have been widely used for an underwater microactuator, cargo displacement assembly, and drug delivery. This study explores a method for underwater cargo transportation using sliding bubbles as a vehicle with directionally guided superhydrophobic wires. By exploitation of the adhesion between superhydrophobic surfaces and bubble interfaces, a bubble is able to transport a superhydrophobic O-ring along a superhydrophobic wire, effectively delivering the O-ring to the water surface.

View Article and Find Full Text PDF

Background: Although fractional flow reserve (FFR) is the contemporary standard to detect hemodynamically significant coronary stenosis, it remains underused for the need of pressure wire and hyperemic stimulus. Coronary angiography-derived FFR could break through these barriers. The aim of this study was to assess the feasibility and performance of a novel diagnostic modality deriving FFR from invasive coronary angiography (AccuFFRangio) for coronary physiological assessment.

View Article and Find Full Text PDF

Background: Several methods for measuring IMR derived from angiography have been developed. AngioIMR is a novel method for the assessment of angiography-derived IMR with no requirement for a wire and hyperemia. The prognostic value of AngioIMR is unknown in STEMI patients.

View Article and Find Full Text PDF

Hibernation enhances contractile responses of basilar artery in ground squirrels: The role of rho-kinase and NO.

Comp Biochem Physiol A Mol Integr Physiol

December 2024

Department of Human and Animal Physiology, Faculty of Biology, M.V. Lomonosov Moscow State University, Moscow, Russia; Department of Physiology and Pathology, Faculty of Basic Medicine, M.V. Lomonosov Moscow State University, Moscow, Russia.

Hibernation is accompanied by dramatic decrease of blood flow in many organs due to the increase of their vascular resistances. We compared the responses of mesenteric, renal, and cerebral proximal resistance arteries in summer active (SA) and winter hibernating (WH) ground squirrels and studied the signaling pathways of Rho-kinase and NO. Wire myography and Western blotting were used to assess the arterial responses and protein abundances.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!