Unlabelled: Interest in mobile-bearing total ankle arthroplasty has increased in recent years. Clinical studies show favorable but varying results, with survival rates between 70% and 90% at 10-year followup. Design-specific differences in early migration patterns might explain variations in the results and modes of failure. Using radiostereometric analysis (RSA) we prospectively followed 15 RA patients with a cementless mobile-bearing total ankle arthroplasty. The American Orthopaedic Foot and Ankle Society ankle score and radiostereometric radiographs were evaluated immediately postoperatively, 6 weeks postoperatively, 3 months, 6 months, and 12 months postoperatively and yearly thereafter. The postoperative clinical results improved. We observed increased migration of the tibial component during the first 3 months, but this stabilized by the 6 month followup. The mean lateral-medial migration was 0.8 mm, distal-proximal migration was 0.9 mm, and posteroanterior migration was -0.5 mm. The latter implies the total resultant migration was in anterior and valgus tilting of this tibial component. However, migration stabilized at 6 months postoperatively. We believe the surgical technique (anterior cortical window for placement) and the method of tibial fixation likely explain this migration.

Level Of Evidence: Therapeutic, Level IV. See Guidelines for Authors for a complete description of levels of evidence.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.blo.0000224013.57916.24DOI Listing

Publication Analysis

Top Keywords

tibial component
12
total ankle
12
early migration
8
migration tibial
8
mobile-bearing total
8
ankle arthroplasty
8
months months
8
months postoperatively
8
migration
7
ankle
5

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!