MR colonoscopy at 3.0 T: comparison with 1.5 T in vivo and a colon model.

Clin Imaging

Klinik für Strahlenheilkunde, Campus Virchow, Charité University Medical Center, Augustenburger Platz 1, 13353 Berlin, Germany.

Published: November 2006

Purpose: Retrospectively, magnetic resonance (MR) colonography images obtained from a colon model and in routine examinations of patients screened for polyps were compared in terms of whether, and to what degree, image quality improved at a higher field strength of 3.0 T compared to 1.5 T.

Materials And Methods: One hundred twenty-eight MR colonography images from 40 patients, of whom 20 had each been scanned at 1.5 and 3.0 T, respectively, using a four-element phased-array torso coil, were compared. At both field strengths, imaging included T1-weighted fat-suppressed spoiled gradient-echo (T1-fs-GE), T2/T1-weighted fast imaging employing steady-state acquisition (FIESTA), and T2-weighted single-shot fast spin-echo (T2-SSFSE), with breath-hold technique. Using receiver operating characteristic analysis performed by seven readers, the three types of images from the colon model and from 20 patients each at 1.5 and 3.0 T were compared. While a time window of 20 s was allowed for picture assessment in a chance-generated succession of images on a monitor, image quality was rated with a score of 1-5 (1=very good; 5=very bad). Statistical significance was calculated with Mann-Whitney U test.

Results: At both field strengths, T2-SSFSE images received the best ratings, followed by FIESTA images (P=.001). Although, overall, the 3.0-T images obtained scores worse than those of the 1.5-T images, a better detection of phantom polyps was noted in the colon model (P=.001).

Conclusion: Although MR colonography with the breath-hold technique using the same four-element phased-array coil at 3.0 and 1.5 T does not perform better at a higher field strength in general, an improved detection of small polyps may be obtained.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.clinimag.2006.02.001DOI Listing

Publication Analysis

Top Keywords

colon model
16
images
8
colonography images
8
images colon
8
image quality
8
higher field
8
field strength
8
four-element phased-array
8
field strengths
8
breath-hold technique
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!