[Precision and accuracy of "a pocket" pulse oximeter in Mexico City].

Rev Invest Clin

Clínica de Suefio, Instituto Nacional de Enfermedades Respiratorias, Tlalpan, DF México.

Published: August 2006

Background: Pulse oximeters are frequently used in the clinical practice and we must known their precision and accuracy. The objective was to evaluate the precision and accuracy of a "pocket" pulse oximeter at an altitude of 2,240 m above sea level.

Methods: We tested miniature pulse oximeters (Onyx 9,500, Nonin Finger Pulse Oximeter) in 96 patients sent to the pulmonary laboratory for an arterial blood sample. Patients were tested with 5 pulse oximeters placed in each of the fingers of the hand oposite to that used for the arterial puncture. The gold standard was the oxygen saturation of the arterial blood sample.

Results: Blood samples had SaO2 of 87.2 +/- 11.0 (between 42.2 and 97.9%). Pulse oximeters had a mean error of 0.28 +/- 3.1%. SaO2 = (1.204 x SpO2) - 17.45966 (r = 0.92, p < 0.0001). Intraclass correlation coefficient between each of five pulse oximeters against the arterial blood standard ranged between 0.87 and 0.99. HbCO (2.4 +/- 0.6) did not affect the accuracy.

Conclusions: The miniature oximeter Nonin is precise and accurate at 2,240 m of altitude. The observed levels of HbCO did not affect the performance of the equipment. The oximeter good performance, small size and low cost enhances its clinical usefulness.

Download full-text PDF

Source

Publication Analysis

Top Keywords

pulse oximeters
20
pulse oximeter
12
arterial blood
12
pulse
8
precision accuracy
8
oximeter
5
oximeters
5
[precision accuracy
4
accuracy pocket"
4
pocket" pulse
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!