[Clinical evaluation of electronic dental analgesia during ultrasonic scaling].

Zhonghua Kou Qiang Yi Xue Za Zhi

Department of Periodontology, Peking Univerisity School and Hospital of Stomatology, Beijing 100081, China.

Published: April 2006

Objective: To evaluate the effect of electronic dental analgesia (EDA) device--Team Up Duo during ultrasonic scaling.

Methods: The clinical trial was conducted as a randomized double-blind split-mouth design. Eighty-one patients requiring ultrasonic scaling were selected as subjects. Patients received treatment using the EDA device either in an active or an inactive state in paired quadrants. The patients scored their pain/discomfort levels with visual analog scale (VAS) after each procedure and filled a questionnaire.

Results: In questionnaire, 79.0% patients felt EDA could relieve pain/discomfort during ultrasonic scaling. But the double-blind test showed that Team Up Duo was effective in 46.9% patients overall. The effective rate was 55.1% and 34.4% in groups with or without attachment loss, respectively. In CP group, the success rate of EDA was not statistically significantly different between the groups according to sex, age or site.

Conclusions: When using Team Up Duo during ultrasonic scaling, most patients subjectively believed that it was effective for controlling pain. In fact, this EDA device was effective in partial patients.

Download full-text PDF

Source

Publication Analysis

Top Keywords

ultrasonic scaling
12
electronic dental
8
dental analgesia
8
duo ultrasonic
8
eda device
8
team duo
8
patients
7
ultrasonic
5
eda
5
[clinical evaluation
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!