Skin entrance radiation dose in an interventional radiology procedure.

Health Phys

Department of Radiological Sciences, University of Oklahoma Health Sciences Center, Oklahoma City, OK 73190, USA.

Published: July 2006

Monitoring of skin entrance radiation exposure in lengthy interventional procedures has been recommended because of the potential for skin injury. Fluoroscopy duration and dose-area product (DAP) are readily available real-time measurements. It would be of interest to study the correlation of these parameters and skin entrance radiation. Twenty neurological interventional procedures performed through the aortic arch were monitored. Two pieces of GafChromic XR Type R film were placed between the patient and the examination table. An observer recorded the fluoroscopy duration and DAP for each phase of the procedure. Each film was scanned post-procedure in RBG mode, and then the image was analyzed for peak skin entrance radiation dose (in air kerma). All DAP values were corrected according to a calibration with an ion chamber. With the DAP values for the respective phases of a procedure, the effective dose in a Reference Man was calculated. For these twenty cases, the means and standard deviations were 17.2+/-6.4 min for x ray on-time, 256+/-65 Gy cm (-2) for DAP, 94+/-34 cGy for peak skin entrance dose in air kerma, and 19.2+/-5.0 mSv for effective dose, respectively. The peak skin entrance dose was correlated to fluoroscopy duration, DAP, and effective dose with the r(2)-values of 0.48, 0.46, and 0.09, respectively. The correlation with DAP or fluoroscopy duration was not sufficiently strong to infer skin entrance dose from either of these parameters. Therefore, skin entrance dose should be determined directly.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.HP.0000198784.94210.1fDOI Listing

Publication Analysis

Top Keywords

skin entrance
32
entrance radiation
16
fluoroscopy duration
16
entrance dose
16
peak skin
12
effective dose
12
skin
9
dose
9
radiation dose
8
interventional procedures
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!