Background: It is important that serological assays detect antibodies to human immunodeficiency virus (HIV) in all infected individuals, including those infected with less prevalent, more diverse subtypes.

Methods: Performance of the ADVIA Centaur HIV 1/O/2 Enhanced (EHIV) Assay was tested on 1344 samples from HIV-positive subjects, 6061 samples from groups at low-risk for HIV infection, and 1042 samples from groups at high-risk for HIV-1 and HIV-2 infection. Results were compared with those of an FDA-approved predicate assay.

Results: The ADVIA Centaur EHIV Assay showed good precision with a diagnostic specificity of 99.9% and diagnostic sensitivity of 100%. HIV seroconversion was detected earlier in 6 panels, at the same time in 13 panels and later in only 1 of the panels when compared to the predicate assay, thereby narrowing the window period between infection and antibody detection. Of clinical significance, a blood donor sample that was indeterminate by HIV-1 Western blot and non-reactive by the predicate assay was repeatedly reactive in the ADVIA Centaur Assay and confirmed as positive by HIV-2 immunoblot.

Conclusions: The ADVIA Centaur EHIV Assay is useful as an aid in the diagnosis of individuals infected with HIV-1 and/or HIV-2.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.cca.2006.04.017DOI Listing

Publication Analysis

Top Keywords

advia centaur
20
ehiv assay
16
centaur hiv
8
hiv 1/o/2
8
1/o/2 enhanced
8
enhanced ehiv
8
samples groups
8
centaur ehiv
8
predicate assay
8
assay
7

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!