Analysis of results related to good and bad outcomes of Intacs implantation for keratoconus correction.

J Cataract Refract Surg

Instituto Oftalmológico de Alicante, Vissum (Department of Cornea and Refractive Surgery), Miguel Hernández University, Medical School, Alicante, Spain.

Published: May 2006

Purpose: To evaluate factors influencing good and poor outcomes after intracorneal ring segment implantation to correct keratoconus.

Setting: Instituto Oftalmológico de Alicante (Vissum Corporation), University of Miguel Hernandez, Alicante, Spain.

Methods: A retrospective comparative study of the visual outcome, 25 eyes of 21 patients were implanted after intracorneal ring segments (Intacs, Addition Technology) as a method to correct keratoconus. Outcomes were evaluated 6 months after implantation based on the visual results. Group A comprised 20 eyes that gained 3 lines or more of best spectacle-corrected visual acuity (BSCVA) to correct keratoconus, and Group B comprised 5 eyes that lost 1 line of BSCVA.

Results: In Group A, the preoperative mean BSCVA was 0.43 (20/50) and the postoperative mean BSCVA, 0.82 (20/20) (P< or =.0001). In Group B, the preoperative mean BSCVA was 0.36 (20/63) and the postoperative mean BSCVA, 0.24 (20/80) (P< or =.02). The preoperative mean sphere power was 3.90 diopters (D) greater in Group B than in Group A (P< or =.007); the mean spherical equivalent was 4.36 D greater (P< or =.007), and the mean K-reading was 6.91 D greater (P< or =.001). The postoperative reduction in spherical dioptric power was 2.11 D (P< or =.027) and the mean spherical equivalent, 2.81 D (P< or =.003), significant only in Group A. The reduction in mean central K-readings was 4.30 D (P< or =.0001) in Group A and 6.19 D (P< or =.014) in Group B. The correlation of the preoperative mean K value to the postoperative BSCVA was significant at the .01 level by the Pearson test (-0.67). In Group A, all eyes (100%) had a mean K-reading < or =53 D. In Group B, 4 eyes (80%) had a mean K-reading > or =55 D.

Conclusion: Intacs provided better results in visual acuity and corneal topography quality and significantly reduced the spherical equivalent in eyes with keratoconus with relatively low mean K values (< or =53 D) and a relatively low spherical equivalent. In advanced keratoconus (mean K-reading > or =55 D), poor results can be anticipated.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jcrs.2006.02.012DOI Listing

Publication Analysis

Top Keywords

spherical equivalent
16
postoperative bscva
12
group
11
intracorneal ring
8
correct keratoconus
8
group comprised
8
comprised eyes
8
visual acuity
8
group preoperative
8
preoperative bscva
8

Similar Publications

Purpose: To evaluate the clinical outcomes of a new FDA approved toric aspheric hydrophobic acrylic intraocular lens.

Patients And Methods: This is a single surgeon, single-arm, on-label, non-randomized, prospective observational study. Thirty patients underwent bilateral cataract surgery (60 eyes) with placement of a Clareon™ Toric IOL in each eye.

View Article and Find Full Text PDF

Stability of refractive outcomes after hyperopic LASIK with and without Mitomycin C application: a randomized controlled trial.

Sci Rep

January 2025

Department of Ophthalmology, Medical Faculty, University Hospital Düsseldorf, Heinrich Heine University, Moorenstraße 5, 40225, Dusseldorf, Germany.

To assess the efficacy, safety, and stability of refractive outcomes in hyperopic Laser-Assisted in Situ Keratomileusis (LASIK) with and without the application of Mitomycin C (MMC). This randomized, parallel group, controlled multicenter trial included 140 hyperopic eyes. The participants were randomly assigned to two groups: one receiving LASIK with mitomycin C (MMC) (n = 70) and the other receiving LASIK without MMC (n = 70).

View Article and Find Full Text PDF

Compartmental analysis of retinal vascular parameters and thickness in myopic eyes using SS-OCTA.

Front Med (Lausanne)

December 2024

Chongqing Key Laboratory of Prevention and Treatment on Major Blinding Diseases, Chongqing Eye Institute, Chongqing Branch (Municipality Division) of National Clinical Research Center for Ocular Diseases, The First Affiliated Hospital of Chongqing Medical University, Chongqing, China.

Background: This study aimed to comprehensively explore the thickness and topographic distributions of retinal vessel alterations of different myopic eyes by using swept-source OCT angiography (SS-OCTA).

Methods: One hundred myopes were included in this observational cross-sectional study. All participants underwent a series of ocular examinations of biometrical parameters, including spherical equivalent refraction (SER), axial length (AL), intraocular pressure (IOP), curvature radius (CR), and others.

View Article and Find Full Text PDF

Visual Performance Correlation with Corneal Aberrometric Profile and Pupil Size After Implantation of a Trifocal Hydrophobic IOL.

Clin Ophthalmol

December 2024

Département de Recherche Clinique, Institut Ophtalmologique de l'Ouest (IOO) Jules Verne, Nantes, France.

Purpose: To evaluate clinical outcomes obtained after cataract surgery involving the implantation of a trifocal hydrophobic intraocular lens (IOL) and to determine if pupil size and the corneal aberrometric profile correlate to visual acuity at different distances.

Methods: 49 patients (98 eyes) underwent bilateral cataract surgery with the placement of FineVision HP IOLs for presbyopia and were assessed at 1- and 3- to 6-months post-surgery. Postoperatively, refraction, monocular and binocular uncorrected and corrected distance visual acuity (UDVA and CDVA), uncorrected intermediate visual acuity (UIVA), uncorrected near visual acuity (UNVA) and the binocular defocus curve were measured.

View Article and Find Full Text PDF

Background: To evaluate the 6-year physiological rates-of-change in ganglion cell inner plexiform layer (GCIPL) and retinal nerve fibre layer (RNFL) thickness measured with optical coherence tomography.

Methods: We included 2202 out of 2661 subjects from the population-based Singapore Chinese Eye Study who returned for follow-up 6 years after baseline examination (follow-up rate 87.7%).

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!