Objectives: Our purpose was to investigate the diagnostic reliability of multi-detector computed tomography (MDCT) in assessing in-stent stenosis compared to digital angiography (DA) in small children.

Background: Little is known about the feasibility of using MDCT to assess stents placed to treat children with congenital heart disease (CHD).

Methods: Twenty-two children (median age [range], 2(3/4) [(1/2) to 12] years) with 42 transcatheter placed stents (median diameter: 7.2 [3.4-16.3] mm) in the pulmonary arteries (n = 36), aorta (2), PDA (1), and SVC (3) underwent both MDCT and DA due to suspected hemodynamic problems.

Results: Independent "blinded" observers were able to measure stent and minimal luminal diameters in 115 out of 124 (93%) stent segments on MDCT and DA. The interobserver variability was low (mean difference: 0.5, SD 0.8 mm) with high correlation (r = 0.97; P < .0001). The percent stenosis by MDCT correlated well with DA (r = 0.89, P < .0001; mean error 2.7, SD 10.4%). For all grades of stenosis, the sensitivity and specificity for MDCT were 58% and 97%, respectively. At a threshold of approximately > or =20% stenosis sensitivity became >98%. All stent associated complications [fracture (4), vascular narrowings (11)] were diagnosed by MDCT. As the stent diameter increased, there was significantly reduced variability between MDCT and DA for in-stent stenosis (P < .0001).

Conclusion: In small children, MDCT is a feasible and promising method for assessing stent associated complications in the treatment of CHD. Cardiac surgeons and interventional cardiologists might rely on this imaging modality to plan specific interventions more precisely and to assess the results upon follow up.

Download full-text PDF

Source
http://dx.doi.org/10.1002/ccd.20760DOI Listing

Publication Analysis

Top Keywords

in-stent stenosis
12
mdct
9
small children
8
children congenital
8
congenital heart
8
heart disease
8
multi-detector computed
8
computed tomography
8
stenosis sensitivity
8
stent associated
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!