pH effect on dynamic coating for capillary electrophoresis of DNA.

Anal Bioanal Chem

College of Chemistry and Molecular Sciences, Wuhan University, Wuhan 430072, PR China.

Published: June 2006

A buffer consisting of tris(hydroxymethyl)aminomethane, 2-(N-moropholino)ethanesulfonic acid (Mes) and EDTA with constant ion strength was used to investigate the effect of buffer pH on the dynamic coating behavior of poly(N-isopropylacrylamide) (PNIPAM) for DNA separation. The atomic force microscopy (AFM) image illustrated that PNIPAM in lower-pH buffer was much more efficient in covering a silica wafer than that in higher-pH buffer. The coating performance of PNIPAM was also quantitatively analyzed by Fourier transform IR attenuated total reflectance spectroscopy and by measuring the electroosmotic flow (EOF). These results indicated that the stability of the dynamic coating was dependent on the pH of the sieving matrix and was improved by reducing the pH to the weak-acid range. The lower pH of the sieving buffer may induce the polymer more efficiently to adsorb on the capillary wall to suppress EOF and DNA-capillary wall interaction for DNA separation. The enhanced dynamic coating capacity of PNIPAM in lower-pH buffer may be attributed to the hydrogen bonds between the hydroxyl groups of the silica surface and the oxygen atom of the carbonyl groups of PNIPAM.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00216-006-0407-4DOI Listing

Publication Analysis

Top Keywords

dynamic coating
16
dna separation
8
pnipam lower-ph
8
lower-ph buffer
8
buffer
6
pnipam
5
dynamic
4
coating capillary
4
capillary electrophoresis
4
electrophoresis dna
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!