Background: Atrial defibrillation can be achieved with a conventional dual-coil, active pectoral implantable cardioverter-defibrillator (ICD) lead system. Shocking vectors that incorporate an additional electrode in the CS have been used, but it is unclear if they improve atrial DFTs.
Objective: The objective of this prospective, randomized study was to determine if a coronary sinus (CS) electrode reduces atrial defibrillation thresholds (DFTs).
Methods: This was a prospective study of 36 patients undergoing initial ICD implant for standard indications. A defibrillation lead with superior vena cava (SVC) and right ventricular (RV) shocking coils was implanted in the RV. An active can emulator (Can) was placed in a pre-pectoral pocket. A lead with a 4 cm long shocking coil was placed in the CS. Atrial DFTs were determined in the following 3 shocking configurations in each patient, with the order of testing randomized: RV --> SVC + Can (Ventricular Triad), distal CS --> SVC + Can (Distal Atrial Triad), and proximal CS --> SVC + Can (Proximal Atrial Triad).
Results: The Proximal and Distal Atrial Triad configurations were both associated with significant reductions in peak current (p < 0.01), but this effect was offset by significant increases in shock impedance (p < 0.01), resulting in no net change in the peak voltage or DFT energy in comparison to the Ventricular Triad configuration (Ventricular Triad: 4.9 +/- 6.6 J, Proximal Atrial Triad: 3.3 +/- 4.1J, Distal Atrial Triad: 4.4 +/- 6.7 J, p > 0.2).
Conclusion: Shocking vectors that incorporate a CS coil do not significantly improve atrial defibrillation efficacy. Since the Ventricular Triad shocking pathway provides reliable atrial and ventricular defibrillation, this configuration should be preferred for combined atrial and ventricular ICDs.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1016/j.hrthm.2006.02.1029 | DOI Listing |
JACC Adv
December 2024
Division of Cardiology, The Johns Hopkins Hospital, Baltimore, Maryland, USA.
This state-of-the-art review examines disparities in the diagnosis, management, and outcomes of cardiac arrhythmias globally. These arrhythmias include atrial fibrillation, ventricular tachyarrhythmias underlying sudden cardiac death, and bradyarrhythmias associated with sinus node and atrioventricular node disease. Arrhythmias in low- and middle-income countries often result in higher mortality rates due to complex and poorly documented risk factors, lack of clinical expertise among health care personnel, lack of sufficient infrastructure, and challenges in access to care.
View Article and Find Full Text PDFCardiol Rev
January 2025
Department of Cardiology, Detroit Medical Center, Detroit, MI.
Congenital heart disease (CHD) is the most common congenital anomaly in newborns. Advances in catheter and surgical techniques led to the majority of these patients surviving into adulthood, leading to evolving challenges due to the emergence of long-term complications such as arrhythmias. Interventional electrophysiology (EP) has had remarkable advances over the last few decades, and various techniques and devices have been explored to treat adult patients with CHD.
View Article and Find Full Text PDFJ Int Med Res
January 2025
Department of Anesthesiology, Shaanxi Provincial People's Hospital, Xi'an, Shaanxi, China.
Monopolar electrocautery is usually a safe and effective technique used in laparoscopic cholecystectomy and bile duct surgery, but it may lead to adverse consequences, even ventricular fibrillation (VF). Amiodarone is an effective antiarrhythmic drug commonly used in practice to treat ventricular and atrial arrhythmias, but it may induce tachyarrhythmia or even VF. We report a case of VF occurring twice during cholecystectomy.
View Article and Find Full Text PDFCureus
December 2024
Cardiology, Quillen College of Medicine, East Tennessee State University, Johnson City, USA.
Cardiac implantable electronic devices (CIEDs), including pacemakers, implantable cardiac defibrillators (ICD), and cardiac resynchronization therapy (CRT) devices, regulate heart rate and rhythm in patients with cardiac conditions. With an aging population, CIED-related complications, especially pacemaker pocket infections, are rising. Risk factors include frailty, older age, and superficial device fixation, while risk mitigation involves larger pocket sizes, submuscular fixation, and absorbable antibacterial envelopes.
View Article and Find Full Text PDFJ Mol Cell Cardiol Plus
June 2024
Nantes Université, CNRS, INSERM, l'institut du thorax, F-44000 Nantes, France.
Heart failure (HF) increases the risk of developing atrial fibrillation (AF), leading to increased morbidity and mortality. Therefore, better prediction of this risk may improve treatment strategies. Although several predictors based on clinical data have been developed, the establishment of a transcriptome-based predictor of AF incidence in HF has proven to be more problematic.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!