Background: The purpose of the European PROCAR Trial was to evaluate the safety and performance of the Protégé stent in the treatment of common and/or internal carotid artery stenoses with adjunctive use of a filter embolic protection device.
Method: The Protégé GPS stent is a self-expanding Nitinol stent system mounted on a 6 Fr 0.018'' (6-9 mm stent) or 7 Fr 0.035'' (10 mm stent) over-the-wire delivery system. Study patient assessments were conducted at baseline, periprocedure, discharge, and 1 and 6 months postprocedure. A total of 77 patients have been enrolled in the trial.
Results: In the 77 lesions treated (31 symptomatic, 46 asymptomatic), the procedure was technically successful in 76 (99%), with an average residual stenosis of less than 30%. One procedure failed because the embolic protection device could not be retrieved and the patient was sent to surgery. Within 30 days, there were four (5.2%) major adverse neurological events (MANEs). Three of the MANEs were major strokes (3.9%), one a minor stroke. The fifth MANE occurred prior to the 6-month follow-up visit; this patient had a major stroke 75 days after the procedure and died 36 days later. One additional death occurred because of urosepsis.
Conclusions: The PROCAR trial shows that the Protégé stent with adjuvant use of a filter embolic protection device satisfies safety and performance criteria for the treatment of carotid artery stenosis. The incidence of MANEs for the Protégé stent is comparable to the incidence of these events in other recent carotid stent studies and standard carotid endarterectomy (CEA).
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1111/j.1540-8183.2006.00132.x | DOI Listing |
CVIR Endovasc
January 2025
Department of Radiology, Section of Vascular and Interventional Radiology, University of Washington, 1959 Northeast Pacific Street, Seattle, WA, 98195, USA.
Purpose: To evaluate access site adverse events following ClotTriever-mediated large-bore mechanical thrombectomy via small upper extremity deep veins (< 6-mm).
Materials And Methods: Twenty patients, including 24 upper extremity venous access sites, underwent ClotTriever-mediated large-bore thrombectomy of the upper extremity and thoracic central veins for symptomatic deep vein obstruction unresponsive to anticoagulation. Patients without follow-up venous duplex examinations (n = 3) were excluded.
JACC Cardiovasc Interv
December 2024
Harrington Heart and Vascular Institute, University Hospitals, Cleveland, Ohio, USA.
JACC Cardiovasc Interv
December 2024
Department for Angiology, Brandenburg Medical School Theodor Fontane, Campus Clinic Brandenburg, Center for Internal Medicine I, Berlin, Germany; Department of Angiology, Sankt-Gertrauden-Krankenhaus, Berlin, Germany.
Background: Several randomized clinical trials have shown that the composite endpoint of death, stroke, and myocardial infarction (MI) is equivalent between carotid artery stenting and carotid endarterectomy. However, the risk of minor stroke has been consistently higher with carotid artery stenting.
Objectives: The authors sought to evaluate the safety and effectiveness of a novel carotid stent system comprised of a stent, an adjustable integrated embolic filter and a postdilation balloon, in patients at elevated risk for adverse events from carotid endarterectomy.
J Int Med Res
January 2025
Department of Anesthesiology, Sir Run Run Shaw Hospital, School of Medicine, Zhejiang University, Hangzhou, Zhejiang, China.
Expiratory central airway collapse is a degenerative tracheobronchial disease that is often overlooked because of its nonspecific clinical features. A man was admitted for evaluation of tracheal nodules. Following bronchoscopic biopsy, a significant increase in airway pressure occurred during anesthesia recovery.
View Article and Find Full Text PDFJ Clin Med
January 2025
Interventional Radiology Department, Adana Dr. Turgut Noyan Application and Research Center, Başkent University, 01250 Adana, Turkey.
: The objective of the current research is to assess the benefits that come with antegrade ureteral stenting coupled with imaging techniques in children and also provide a new classification based on the ureter's morphological elements. : Between 2011 and 2024, 107 antegrade stent placement procedures performed in 71 pediatric patients aged 0-12 years who could not undergo retrograde double-J stent placement were retrospectively analyzed. According to the morphologic structure of the ureter, four categories were classified as normal, straight/slightly angled, S-shaped, and spiral-shaped.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!