Objective: Current guidelines recommend that only validated devices for blood pressure measurement should be used. This study presents the validation results of the Microlife BPA100 Plus monitor for self-home blood pressure measurement, according to the European Society of Hypertension International Protocol.
Methods: Fifteen study participants were included in phase 1 and an additional 18 in phase 2 (total 33). Simultaneous blood pressure measurements were taken by two trained observers (Y tube connected mercury sphygmomanometers) four times, sequentially with three measurements taken using the tested device. Absolute differences between observer and device blood pressure were classified into three zones (within 5, 10 and 15 mmHg). The number of readings with a difference within 5 mmHg was calculated for each individual.
Results: In phase 1, the device produced 32, 42 and 43 measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure and 31, 45 and 45 for diastolic blood pressure. In phase 2.1, the device produced 71, 87 and 96 measurements within 5, 10 and 15 mmHg, respectively, for systolic blood pressure, and 71, 98 and 99 measurements for diastolic blood pressure. In phase 2.2, 26 participants had at least two of their differences within 5 mmHg and three participants had no differences within 5 mmHg for systolic blood pressure, whereas for diastolic blood pressure 24 and three participants, respectively. Mean systolic blood pressure differences were -2.0+/-6.0 mmHg and diastolic blood pressure differences were -3.1+/-4.1 mmHg.
Conclusions: The Microlife BPA100 Plus device for self-home blood pressure measurement comfortably passes the validation requirements of the International Protocol and therefore can be recommended for clinical use in the adult population.
Download full-text PDF |
Source |
---|---|
http://dx.doi.org/10.1097/01.mbp.0000209071.84965.bf | DOI Listing |
Sci Rep
December 2024
School of Gongli Hospital Medical Technology, University of Shanghai for Science and Technology, No. 516 Jungong Road, Yangpu District, Shanghai, 200093, China.
Transjugular intrahepatic portosystemic shunt (TIPS) is a widely used surgery for portal hypertensive patients, whose potential postoperative complications are closely related to the hemodynamic condition of the portal venous system. The selection of shunt position in the surgery may affect the postoperative hemodynamics; however, it is difficult for clinical studies to investigate the influence. Therefore, this study aims to employ the computational model simulating TIPS to compare the hemodynamic differences resulting from different shunt positions, and also to investigate the influences of different geometrical model simplification strategies used in the TIPS simulation.
View Article and Find Full Text PDFTrials
December 2024
Department of Cardiology, The Heart Centre, Copenhagen University Hospital Rigshospitalet, Copenhagen, Denmark.
Background: Intermediate-high risk pulmonary embolism (PE) carries a significant risk of hemodynamic deterioration or death. Treatment should balance efficacy in reducing clot burden with the risk of complications, particularly bleeding. Previous studies on high-dose, short-term thrombolysis with alteplase (rtPA) showed a reduced risk of hemodynamic deterioration but no change in mortality and increased bleeding complications.
View Article and Find Full Text PDFBMC Pediatr
December 2024
Department of Pediatric Intensive Care, Faculty of Medicine, Cukurova University, Adana, Turkey.
Background: Albumin, a vital component in regulating human blood oncotic pressure, plays an important role in the prediction of prognosis in pediatric patients.Previous research identified significant differences in serum albumin levels of healthy and critically ill children.
Methods: The present study aims to investigate the correlation between albumin levels measured during pediatric intensive care unit(PICU) admission and clinical outcomes.
Mol Med
December 2024
Department of Otorhinolaryngology/Head and Neck, Sir Run Run Shaw Hospital, Zhejiang University School of Medicine, No.3 East Qingchun Road, Hangzhou, 310020, Zhejiang, China.
Background: Sleep apnea syndrome (SAS) is associated with hypertension and vascular remodeling. Hypoxia-inducible factor-1α (HIF-1α) and the Hippo-YAP pathway are implicated in these processes, but their specific roles remain unclear. This study investigated the HIF-1α/Hippo-YAP pathway in SAS-related hypertension.
View Article and Find Full Text PDFBMC Pregnancy Childbirth
December 2024
Department of Public Health, Department of Anesthesiology, Second Affiliated Hospital of Zhejiang University School of Medicine, Hangzhou, China.
Background: Maternal hemoglobin (Hb) is related to nutritional status, and it widely fluctuates during pregnancy. However, the relationship between Hb and blood pressure (BP) during pregnancy is unclear. This study aimed to estimate the associations between maternal Hb in various trimesters and its changes with BP during pregnancy.
View Article and Find Full Text PDFEnter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!