Prosthetic manufacturers have marketed shock-absorbing pylons (SAPs) for attenuation of injurious loads from foot-ground contact. In this study, we compared a commonly prescribed SAP with a conventional rigid pylon, using a within-subject design (n = 15 unilateral transtibial amputees), to assess effect on gait mechanics, measure transmitted accelerations in situ, and determine functional outcomes using step counts and questionnaires. No differences were found across pylons for self-selected walking speed, prosthetic-side step length, prosthetic-side loading rate and decelerative peak of the vertical ground reaction force, peak pylon acceleration, step count per week, or questionnaire results that examined pylon performance and subjects' pain and fatigue levels. The only statistically significant finding was for the prosthetic-side knee angle at initial contact, where subjects displayed an average of 2.6 degrees more flexion with the rigid pylon than the SAP while walking at a controlled speed (p = 0.004); this result indicates that transtibial amputees are able to modulate the effective stiffness of their residual limb in response to changes in prosthetic component stiffness. The results from the laboratory, field, and subjective outcome measurements suggest that the SAP in this study is as effective as a rigid pylon for unilateral transtibial amputees.

Download full-text PDF

Source
http://dx.doi.org/10.1682/jrrd.2005.02.0034DOI Listing

Publication Analysis

Top Keywords

transtibial amputees
16
rigid pylon
12
laboratory field
8
unilateral transtibial
8
pylon
5
efficacy shock-absorbing
4
shock-absorbing versus
4
rigid
4
versus rigid
4
rigid pylons
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!