A series of in vitro patella damage tests were developed using a knee simulator to investigate electron-beam-irradiated and -melted ultra-high molecular weight polyethylene and conventional ultra-high molecular weight polyethylene at the patellofemoral articulation. Three different simulations were created: (i) normal gait with optimal component alignment, (ii) stair climbing with optimal component alignment, and (iii) stair climbing with 4 degrees of femoral component internal rotation to simulate a component malalignment condition. In the last two simulations, all patellae were artificially aged. In normal gait, the unaged conventional and highly crosslinked patellae demonstrated similar behavior. In both stair climbing tests, unlike the aged highly crosslinked components, the aged conventional patellae developed cracks by 2 million cycles. These results demonstrate the potential advantage of highly crosslinked polyethylene for the patella.

Download full-text PDF

Source
http://dx.doi.org/10.1002/jbm.b.30538DOI Listing

Publication Analysis

Top Keywords

highly crosslinked
16
stair climbing
12
conventional highly
8
ultra-high molecular
8
molecular weight
8
weight polyethylene
8
normal gait
8
optimal component
8
component alignment
8
comparison conventional
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!