Effects of rounding errors on postmortem temperature measurements caused by thermometer resolution.

Int J Legal Med

Division of Forensic Medicine, Department of Public Health and Forensic Medicine, Tohoku University School of Medicine, Sendai 980-8575, Japan.

Published: July 2007

Beginning 7 h after death, a datalogger was used to measure the temperature in the external auditory canal of an adult male body placed in a refrigerated room. The sequence of measured values approximated a single exponential function with a correlation coefficient of 0.998475. This suggests that the starting time of body cooling in the refrigerated room under constant temperature can be calculated with less error using any two data points recorded by the datalogger. However, the results of such calculations varied widely and longer postmortem intervals demonstrated greater calculation errors. Periodic errors also appeared. Mathematical simulations showed that this variation was caused by rounding errors, which represent the difference between the thermometer readings and the true temperature. The resolution of the thermometer was 0.1 degrees C, a normal specification; however, even this led to noticeable rounding errors. Therefore, significant errors may influence postmortem interval estimations using other body temperatures. When body temperatures are used to determine the time of death, a method that minimizes rounding errors should be considered.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00414-006-0088-8DOI Listing

Publication Analysis

Top Keywords

rounding errors
16
refrigerated room
8
body temperatures
8
errors
7
effects rounding
4
errors postmortem
4
temperature
4
postmortem temperature
4
temperature measurements
4
measurements caused
4

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!