Purpose: The aim was to investigate the feasibility and image quality of prospective respiratory gating for 3-D computed tomography (CT) of the lung.

Material And Methods: Eight anesthetized pigs underwent prospectively gated multidetector computed tomography using 2 devices: a charge-coupled device (CCD) camera and a laser sensor. The output signal of both gating devices was connected to the scanner instead of ECG unit. Inspiratory and expiratory images were obtained during "free-breathing" and analyzed in MPR mode for sharpness of bronchi, diaphragm and lung using a 4-point-score (1, excellent to 4, severe artifacts).

Results: The CCD camera worked in all animals. Using the laser sensor, only 50% of expiratory scans could be acquired. All acquired images showed excellent sharpness (CCD camera vs. laser sensor) for trachea (1.1 +/- 0.3 vs. 1.3 +/- 0.5), bronchi (1.4 +/- 0.7 vs. 1.8 +/- 0.6), lung fissures (1.0 vs. 1.1 +/- 0.3), and lung parenchyma (1.0 +/- 0.2 vs. 1.4 +/- 0.6), and minor to major artifacts for diaphragm (1.5 +/- 0.8 vs. 2.0 +/- 1.0, P < 0.05) and pericardial lung structures (1.9 +/- 0.7 vs. 2.3 +/- 0.5).

Conclusion: High image quality for inspiratory and expiratory scans was achieved by free-breathing 3-D CT of the lung using noncontact prospective respiratory gating.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.rli.0000208926.98693.b6DOI Listing

Publication Analysis

Top Keywords

+/- +/-
20
laser sensor
16
computed tomography
12
prospective respiratory
12
respiratory gating
12
camera laser
12
ccd camera
12
+/-
11
charge-coupled device
8
image quality
8

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!