Background: Implantable cardioverter defibrillator (ICD) is the only effective therapy in patients with life threatening ventricular arrhythmias. Inappropriate detection and therapy by ICDs are the most common causes of side effects that affect the quality of life in ICD recipients. This study evaluated the incidence and causes of inappropriate detection and therapy by ICDs in patients in our hospital.

Methods: From January 2000 to December 2005, fifty patients who received ICD implantation for ventricular arrhythmias for prevention of sudden cardiac death were evaluated in this study. Each ICD was programmed using clinical arrhythmic and cardiac data of the patient before discharge. Patients were followed up by standard schedule after implantation and all data retrieved from each device were collected and saved for further analysis.

Results: No arrhythmic event was detected in 12/50 (24%) patients during the period of follow-up. Among the remaining patients, 11 (22%) experienced inappropriate detections and therapies during follow-up in this study. ICD detected 383 ventricular tachyarrhythmia (VT) and 108 ventricular fibrillation (VF) episodes and delivered 678 therapies. In VT group, ICD delivered 413 antitachycardiac pacings (ATPs) and 118 shocks, among which 78 ATPs and 9 shocks were initiated by 55/383 (14.3%) inappropriate detections. In VF group ICD delivered 147 shocks, among which 56 shocks were initiated by 28/108 (26.9%) inappropriate detections. Overall, more than 50% of these episodes were caused by atrial fibrillation (AF) with rapid ventricular response, followed by electromagnetic or myopotential interference. In addition, most inappropriate therapies occurred within one year after ICD implantation.

Conclusions: About one fifth of patients experienced ICD inappropriate detection and therapy after implantation. The main cause was AF with rapid ventricular response, followed by electromagnetic or myopotential interference.

Download full-text PDF

Source

Publication Analysis

Top Keywords

inappropriate detection
16
detection therapy
16
inappropriate detections
12
icd
9
incidence inappropriate
8
patients
8
ventricular tachyarrhythmia
8
ventricular arrhythmias
8
therapy icds
8
study icd
8

Similar Publications

From stroke to depression: The need for systematic screening for post-stroke depression.

S Afr J Psychiatr

December 2024

Department of Internal Medicine, Cameroon Baptist Convention Health Services, Bamenda, Cameroon.

Background: Post-stroke depression (PSD) negatively impacts the physical and mental well-being of stroke survivors. However, data on the prevalence and risk factors of PSD in African countries such as Cameroon are limited.

Aim: This study aims to determine the prevalence and factors associated with PSD among stroke survivors at a hospital in Cameroon and inform clinical practice.

View Article and Find Full Text PDF

Compositional data are multivariate data made up of components that sum to a fixed value. Often the data are presented as proportions of a whole, where the value of each component is constrained to be between 0 and 1 and the sum of the components is 1. There are many applications in psychology and other disciplines that yield compositional data sets including Morris water maze experiments, psychological well-being scores, analysis of daily physical activity times, and components of household expenditures.

View Article and Find Full Text PDF

Background: Dosimetric commissioning and quality assurance (QA) for linear accelerators (LINACs) present a significant challenge for clinical physicists due to the high measurement workload and stringent precision standards. This challenge is exacerbated for radiosurgery LINACs because of increased measurement uncertainty and more demanding setup accuracy for small-field beams. Optimizing physicists' effort during beam measurements while ensuring the quality of the measured data is crucial for clinical efficiency and patient safety.

View Article and Find Full Text PDF

Objectives: To develop a scoring system to predict resistance to ceftolozane/tazobactam in Pseudomonas aeruginosa strains isolated from respiratory specimens.

Methods: A case-control study was conducted to evaluate the risk factors associated with resistance to ceftolozane/tazobactam. Patients with P.

View Article and Find Full Text PDF

Unlabelled: Ongoing research in biosensor technologies has led to advanced functional materials for healthcare diagnostics, and bacteriophages (phages), demonstrating exceptional utility due to their high specificity, accuracy, rapid, label-free, and wireless detection capabilities with minimal false-positive results. Phage-based-pathogen-detecting biosensors (PBPDBs) include surface plasmon resonance (SPR) biosensors, magnetoelastic (ME), electrochemical, and quartz crystal microbalance (QCM) biosensors. Commonly used substrates for PBPDBs are gold, silicon, glass, carbon-based materials, magnetic particles, and quantum dots.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!