Background And Study Aims: This study evaluated the effectiveness of the cleaning process, the disinfection process, and a combination of the two in accordance with the new international standard, prEN ISO 15 883.

Materials And Methods: The cleaning process consisted of a 1-min prerinse at 20 degrees C, followed by a 5-min cleaning step at 45 degrees C (with an enzymatic cleaner, 0.5 %), followed by a 1-min interim rinse from 45 degrees C to 55 degrees C. The disinfection process consisted of a 1-min prerinse at 20 degrees C, followed by a 5-min disinfection step at 55 degrees C (with a glutaraldehyde-based disinfectant, 1 %), followed by two final rinses of 1 min each at 55 degrees C. Transparent test pieces were contaminated with a mixture of blood and ENTEROCOCCUS FAECIUM, and were assessed for visible cleanliness and microbial load.

Results: Cleaning alone, disinfection alone, and the combination of the two always led to visible cleanliness of all test pieces. The cleaning process revealed a mean reduction factor of > or = 4.6 (n = 6); the disinfection process revealed a mean reduction factor of > or = 9.0 (n = 6), and the combination of the two was found to reduce the test organism in the WD440 by 9.0 +/- 0.2 log (10) steps (n = 12) and in the AdaptaScope by 9.3 +/- 0.4 log (10) steps (n = 5).

Conclusions: Overall, the entire process was found to be very effective and compatible for reprocessing flexible endoscopes in washer-disinfectors. No visible residual blood was found, despite the use of glutaraldehyde in the disinfection phase. These findings once again emphasize the importance of effective cleaning for the overall results when reprocessing flexible endoscopes.

Download full-text PDF

Source
http://dx.doi.org/10.1055/s-2006-925133DOI Listing

Publication Analysis

Top Keywords

flexible endoscopes
12
cleaning process
12
disinfection process
12
enzymatic cleaner
8
glutaraldehyde-based disinfectant
8
endoscopes washer-disinfectors
8
pren iso
8
process consisted
8
consisted 1-min
8
1-min prerinse
8

Similar Publications

Endobronchial ultrasound-guided transbronchial needle aspiration (EBUS-TBNA) is a cornerstone in minimally invasive thoracic lymph node sampling. In lung cancer staging, precise assessment of lymph node position is crucial for clinical decision-making. This study aimed to demonstrate a new deep learning method to classify thoracic lymph nodes based on their anatomical location using EBUS images.

View Article and Find Full Text PDF

This article comments on the article by Du , who conducted a randomized controlled trial aiming at evaluating the effectiveness of a novel spray flushing system in cleaning flexible endoscopes while minimizing damage to the working channels. We share our perspective on the importance of improving endoscope reprocessing methods. The findings highlight the spray flushing system's capacity to improve cleaning efficacy while minimizing damage, suggesting that it might be important in enhancing endoscope reprocessing procedures.

View Article and Find Full Text PDF

Currently, Unilateral biportal endoscopy is widely used in the surgical treatment of lumbar spinal stenosis. To investigate the feasibility of bilateral synchronous UBE to unilateral laminotomy and bilateral decompression(BS-UBE-ULBD) for treating two-level lumbar spinal stenosis (LSS). Sixty-four patients with two-level lumbar spinal stenosis (LSS) treated with BS-UBE-ULBD from October 2022 to January 2024 were retrospectively analyzed.

View Article and Find Full Text PDF

Background: Foreign body inhalation is rare in older children, often leading to underdiagnosis and delayed treatment. Most cases involve a single foreign body, but instances of multiple foreign bodies are exceedingly uncommon. This report presents a case of an elder child who inhaled two pen caps, emphasizing the need for clinical vigilance and thorough medical history collection.

View Article and Find Full Text PDF

Purpose: To report on in-vivo intrarenal pressure (IRP) during flexible ureteroscopy for treatment of kidney stones while using the novel flexible and navigable suction ureteral access sheath (FANS).

Methods: A retrospective observational analysis was performed for 25 patients undergoing routine flexible ureteroscopy for treatment of renal stones between February 2024 and June 2024 from two centres in Canada. The LithoVue Elite™ ureteroscope (Boston Scientific Corp.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!