Development and validation of a rapid 96-well format based liquid-liquid extraction and liquid chromatography-tandem mass spectrometry analysis method for ondansetron in human plasma.

J Chromatogr B Analyt Technol Biomed Life Sci

Laboratory of Pharmaceutical Analysis and Bioequivalence Services, Department of Pharmaceutical Chemistry, School of Pharmacy, University of Athens, Greece.

Published: May 2006

A rapid and sensitive LC-MS/MS method for the quantification of ondansetron was developed and validated. The plasma samples were treated by a semi-automated liquid-liquid extraction (LLE) in 1.2 mL 96-well format micro-tubes. Ondansetron and the internal standard (IS) granisetron were analyzed by combined reversed phase LC-MS/MS, with positive ion electrospray ionization, using multiple reactions monitoring (MRM). The statistical evaluation for this method reveals excellent linearity, accuracy and precision values for the range of concentrations 0.25-40.0 ng/mL. The proposed method enabled the reliable determination of ondansetron in bioequivalence studies after per os administration of a 4 or 8 mg tablet.

Download full-text PDF

Source
http://dx.doi.org/10.1016/j.jchromb.2006.03.032DOI Listing