The free-field transmitting voltage response at the pressure focus of a spherically focusing transducer was defined and calibrated based on the reciprocity theorem of a free-field spherically focused acoustic wave. The acoustic power, the radiation conductance, and the pressure at the pressure focus were derived and measured accordingly from the transmitting current response on the imaginary mirror symmetric spherical surface of the radiating surface. A miniature hydrophone was calibrated by the self-reciprocity of the spherically focusing source. Comparison results show that the measured acoustic power deviation between the reciprocity method and the radiation force balance method are within +/- 5% for two air-backed focusing transducers at 1.53 MHz and 5.27 MHz, respectively, and the maximum deviation of a hydrophone calibration between the new method and the free-field plane wave reciprocity method is within 1.4 dB in the frequency range from 1 MHz to 2 MHz in experiments.