Background: Plain radiography, subtraction arthrography, nuclear arthrography, and bone scintigraphy are imaging techniques commonly used to identify aseptic femoral component loosening. Controversy exists about the relative utility of these techniques.

Patients And Methods: We evaluated the diagnostic accuracy and interobserver reliability of the four techniques in 78 consecutive patients (mean age 70 years, range 29-88 years) referred for evaluation of their femoral hip prostheses. The standard evaluation protocol consisted of plain radiography followed by subtraction arthrography, nuclear arthrography, and bone scintigraphy. Surgery or the subsequent clinical course of the patient was used as gold standard.

Results: Overall, plain radiography had a sensitivity and specificity of 81 and 74%, respectively. Subtraction arthrography had a sensitivity of 47% and a specificity of 78%. Nuclear arthrography had a sensitivity of 69% and a specificity of 76%, and bone scintigraphy had a sensitivity of 88% with a specificity of 50%.

Conclusion: We found considerable interobserver variability in all four techniques. Multivariate regression analysis revealed that bone scintigraphy and nuclear arthrography together made a significant contribution to the diagnosis when used in combination with plain radiography and are, when plain radiography is inconclusive, useful additional diagnostic techniques for the detection of femoral component loosening.

Download full-text PDF

Source
http://dx.doi.org/10.1007/s00402-006-0120-yDOI Listing

Publication Analysis

Top Keywords

plain radiography
24
nuclear arthrography
20
bone scintigraphy
20
subtraction arthrography
16
radiography subtraction
12
arthrography nuclear
12
arthrography bone
12
femoral component
12
component loosening
12
arthrography
9

Similar Publications

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!