Objective: Assessment of the optimal static preloading of Otologics Middle Ear Transducer (MET) Ossicular Stimulator, when coupled to the incus.

Background: The MET Ossicular Stimulator is a partially implantable electromagnetic middle ear hearing device that transmits vibrations to the ossicular chain. The vibration patterns were measured with laser-Doppler vibrometry.

Study Design: Experimental.

Material: We used three human cadaveric temporal bones (TB) and one MET ossicular stimulator.

Methods: Laser-Doppler vibrometry was used for the selection of TBs. The cochlea was subsequently extirpated from the posterior side to measure the vibrational patterns (VP) of the footplate. Three TBs with different VP were selected based on data obtained from volunteers with normal hearing (n = 110): one TB with a VP larger than +1 SD, one TB with a VP in the range of +/-1 SD, and 1 TB with a VP smaller than -1 SD. Transfer functions were calculated between VP of the measurement points at the coupling rod, umbo, incus, and footplate. The TBs were subsequently defrosted. The MET was implanted and coupled to the ossicular chain. Different coupling loads were measured at the incus, the umbo, and the footplate.

Results: Optimal transfer function between the MET transducer and the oval window was achieved during contact when the coupling rod advanced 0.0625 mm (90 degrees rotation). Additional advances of 0.0625 mm (180 degrees turn = 0.125 mm) resulted in a decreased vibrational amplitude, ranging between 20 and 40 dB below 3 kHz. The lowest linear distortion occurred up to 10 kHz during direct contact without advancing the coupling rod.

Download full-text PDF

Source
http://dx.doi.org/10.1097/01.MLG.0000199889.61483.8FDOI Listing

Publication Analysis

Top Keywords

met ossicular
12
coupling rod
12
middle ear
8
ossicular stimulator
8
ossicular chain
8
met
6
ossicular
5
vitro model
4
model intraoperative
4
intraoperative adjustments
4

Similar Publications

Realizations of fully implantable cochlear implants (FICIs) for providing adequate solution to esthetic concerns and frequent battery replacement have lacked of addressing system level criteria as a complete device. Here, we present a full-custom FICI that considers design of both an implantable sensor for wide range sound sensing and a signal conditioning circuit for electrical stimulation of the auditory nerve. The microelectromechanical system (MEMS)-based acoustic sensor utilizes multiple cantilever beam structures to sense and filter the mechanical vibrations on the ossicular chain.

View Article and Find Full Text PDF

Diagnostic Value of Cone-Beam Computed Tomography in Conductive or Mixed Hearing Loss with Intact Tympanic Membrane.

J Otolaryngol Head Neck Surg

September 2024

Department of Otorhinolaryngology-Head and Neck Surgery, the First Affiliated Hospital of Guangxi Medical University, Nanning, Guangxi, People's Republic of China.

Background: Conductive or mixed hearing loss with an intact tympanic membrane is a group of diseases characterized by similar clinical symptoms. Definitive diagnosis depends on the findings of exploratory tympanic surgery. Cone-beam computed tomography (CBCT) has great potential for middle ear imaging.

View Article and Find Full Text PDF

Aims: In this study, we attempt to compare the pure tone audiometry findings with ossicular chain status intraoperatively in patients of chronic otitis media.

Materials And Methods: 102 patients who presented with COM during a period of one year and met the inclusion criteria were included in the study. All patients underwent preoperative pure tone Audiometry and findings were tabulated.

View Article and Find Full Text PDF

Design of Piezoelectric Dual-Bandwidth Accelerometers for Completely Implantable Auditory Prostheses.

IEEE Sens J

July 2023

Department of Mechanical Engineering and the Department of Biomedical Engineering, University of Michigan, Ann Arbor, MI, 48109, USA.

For the last 20 years, researchers have developed accelerometers to function as ossicular vibration sensors in order to eliminate the external components of hearing aid and cochlear implant systems. To date, no accelerometer has met all of the stringent performance requirements necessary to function in this capacity. In this work, we present an accelerometer design with an equivalent noise floor less than 20 phon equal-loudness-level over a 0.

View Article and Find Full Text PDF
Article Synopsis
  • Endoscopic ossicular chain reconstruction (OCR) in children has shown similar results to traditional methods, but research on its effectiveness for pediatric patients is limited due to unique anatomical and pathological factors.
  • A retrospective study at a tertiary care center reviewed 17 children under 17 years old who underwent endoscopic OCR, focusing on surgical outcomes, hearing improvement, and complications.
  • The typical age of patients was about 11.3 years; most had congenital cholesteatoma, experienced no major complications, and had a long-term follow-up of 2.6 years after surgery.
View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!