Background: The expanding indications for sirolimus-eluting stents (SES) include increasingly complex coronary lesions and populations with clinical profiles markedly different from those of early pivotal controlled studies. The e-Cypher registry monitored the safety and efficacy of SES currently implanted worldwide in daily practice.

Methods And Results: Between April 2002 and September 2005, data were collected on 15,157 patients who underwent implantation of > or =1 SES at 279 medical centers from 41 countries. An independent endpoint review committee adjudicated all reported major adverse cardiovascular events, stent thromboses, and target-vessel revascularizations. Data were managed and analyzed by independent organizations. Predictors of adverse clinical events were identified by regression analysis. The mean age of the sample was 61.7+/-11.4 years; 77.7% were men, and 28.6% were diabetics. A total of 18,295 lesions were treated (20,503 SES) during the index procedure. The cumulative rates of major adverse cardiovascular events were 1.36% at 30 days, 3.38% at 6 months, and 5.80% at 1 year. The rates of acute, subacute, and late stent thrombosis were 0.13%, 0.56%, and 0.19% of patients, respectively, representing a 12-month actuarial incidence of 0.87%. Insulin-dependent diabetes, acute coronary syndrome at presentation, and advanced age were clinical predictors, whereas TIMI flow grade <3 after the index procedure, treatment of multiple lesions, a prominently calcified or totally occluded target lesion, and multivessel disease were the angiographic or procedural predictors of stent thrombosis at 12 months.

Conclusions: This analysis of 1-year data collected by the e-Cypher registry suggests a high degree of safety of SES, with a rate of stent thrombosis similar to that observed in randomized trials.

Download full-text PDF

Source
http://dx.doi.org/10.1161/CIRCULATIONAHA.104.532242DOI Listing

Publication Analysis

Top Keywords

sirolimus-eluting stents
8
e-cypher registry
8
major adverse
8
adverse cardiovascular
8
cardiovascular events
8
safety coronary
4
coronary sirolimus-eluting
4
stents daily
4
clinical
4
daily clinical
4

Similar Publications

Introduction And Objectives: Patients undergoing percutaneous coronary intervention in vessels with moderate-to-severe tortuosity are at higher risk of adverse outcomes, but data are scarce in the era of newer-generation stents. We compared outcomes following percutaneous coronary intervention in vessels with moderate-to-severe tortuosity using a bioresorbable-polymer sirolimus-eluting stent (BP-SES) vs a durable-polymer everolimus-eluting stent.

Methods: A total of 2350 patients from the BIOFLOW II, IV, and V randomized trials were stratified into 2 groups based on target-vessel tortuosity: none-to-mild and moderate-to-severe.

View Article and Find Full Text PDF

Limited evidence exists regarding the long-term outcomes of true versus nontrue coronary bifurcation lesions (CBLs) treated with current-generation drug-eluting stents and intravascular imaging guidance. The Sapporo Cardiovascular Clinic (SCVC) registry was a prospective, single-center, all-comers registry enrolling 1,727 consecutive patients treated with bioresorbable polymer sirolimus-eluting stent under complete imaging guidance. From this registry, 440 patients with CBLs (25.

View Article and Find Full Text PDF

Background: This study evaluated the safety and efficacy of BioMime sirolimus-eluting stent (SES) system, with an ultra-low strut thickness (65 µm), in real-world all-comers population with coronary artery stenosis (CAD).

Methods: This was a post-marketing, multicenter, single-arm, observational clinical registry among patients undergoing intervention for CAD. Patients were clinically followed up at 1, 9, 12, and 24 months after the index percutaneous coronary intervention.

View Article and Find Full Text PDF

Background: The Orsiro and Genoss DES stents are biodegradable polymer drug-eluting stents (DESs) with ultrathin struts.

Objective: To investigate the safety and efficacy of these two ultrathin DESs in real-world practice.

Methods: From a single-center prospective registry, we included 751 and 931 patients treated with the Genoss DES and Orsiro stents, respectively.

View Article and Find Full Text PDF

Background: Although coronary artery involvement in patients with IgG4-related disease (IgG4-RD) is rare, emergency revascularization is recommended for managing acute coronary syndrome. However, coronary aneurysm formation and stent migration after sirolimus-eluting stent implantation have been reported for this disease. Thus, new treatment modalities are warranted for the management of coronary artery disease in this vasculitis.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!