Background: A novel magnetic navigation system (MNS) allowing remote guidance of catheters and guidewires might assist in implantation of left ventricular (LV) pacing leads.

Objective: To assess the feasibility of deploying a LV pacing lead into a coronary sinus (CS) side branch using a magnetically guided wire and of performing the procedure without a CS guiding sheath.

Methods: Twenty-one patients were included in this study. Nine underwent CRT device implantation using a MNS to steer the guidewire (MNS group) while 12 patients were conventionally implanted (control group). In 6 patients in the MNS group, the procedure was performed using a CS guiding sheath. In 3 others, the decision was to perform the procedure without a CS sheath. In these patients the wire was advanced manually, while the external magnets oriented it toward the CS os. In the CS, "vector based" navigation was used to guide the wire to the desired side branch.

Results: In all 9 patients in the MNS group, the target vessel could be successfully engaged by the magnetically guided wire. In 7, the LV lead was lodged in the target vessel. In 2 patients, the LV lead was repositioned in an anterolateral side branch due to instability or inability to engage the vessel with it. Mean total procedure time was 164 +/- 58 minutes (without sheath 229 +/- 52 vs with sheath 132 +/- 26 minutes; P = 0.007). Mean fluoroscopy time was 28 +/- 9 minutes. For control patients, the procedure and fluoroscopy time were similar (144 +/- 41 minutes and 26 +/- 12 minutes, respectively). No major complications occurred.

Conclusion: LV lead implantation can be performed using a remote magnetically steered guidewire. Though the lead could be implanted without a CS guiding sheath, longer procedure times were required.

Download full-text PDF

Source
http://dx.doi.org/10.1111/j.1540-8167.2005.00313.xDOI Listing

Publication Analysis

Top Keywords

+/- minutes
20
side branch
12
mns group
12
left ventricular
8
coronary sinus
8
sinus side
8
magnetic navigation
8
magnetically guided
8
guided wire
8
group patients
8

Similar Publications

Not too much, not too little. Titrating flow rate to minimise inspiratory effort during helmet CPAP: A bench study.

Pulmonology

December 2025

Department of Emergency and Intensive Care, Fondazione IRCCS San Gerardo dei tintori, Monza, Italy.

Background: Non-invasive helmet respiratory support is suitable for several clinical conditions. Continuous-flow helmet CPAP systems equipped with HEPA filters have become popular during the recent Coronavirus pandemic. However, HEPA filters generate an overpressure above the set PEEP.

View Article and Find Full Text PDF

Importance: High-quality colonoscopy reduces the risks of colorectal cancer by increasing the adenoma detection rate. Routine use of an automatic quality control system (AQCS) to assist in colorectal adenoma detection should be considered.

Objective: To evaluate the effect of an AQCS on the adenoma detection rate among colonoscopists who were moderate- and low-level detectors during routine colonoscopy.

View Article and Find Full Text PDF

Pulmovermis cyanovitellosus Coil and Kuntz, 1960 is a species of hemiurid trematode that localizes in the lung of sea snakes, an unusual trait for this group of parasites. Recent molecular phylogenetic studies based on 28S rRNA gene sequences have shown that this species is closely related to members of the genus Lecithochirium Lühe, 1901. This finding is unexpected given that Pulmovermis Coil and Kuntz, 1960 and Lecithochirium are currently classified in different subfamilies of Hemiuridae (Pulmoverminae Sandars, 1961 vs.

View Article and Find Full Text PDF

Objectives: The purpose of this study was to determine the accuracy of the Relative Value Update Committee (RUC) and Centers for Medicare & Medicaid Services current times and work relative value units (wRVUs) for the perioperative work involved in anterior cruciate ligament (ACL) reconstruction by directly timing perioperative tasks as they occur in real time.

Methods: The RUC was contacted to obtain a list of perioperative tasks and the corresponding times allotted for the tasks involved in arthroscopically aided ACL reconstruction (Current Procedural Terminology code 29888). The tasks that occurred both inside and outside the operating room were timed by the attending physician as the event occurred.

View Article and Find Full Text PDF

Want AI Summaries of new PubMed Abstracts delivered to your In-box?

Enter search terms and have AI summaries delivered each week - change queries or unsubscribe any time!